NOTE:
The thread on the test port, and the lower part of the gauge adapters is LEFT-
HANDED. The following procedure details the correct method for mounting devices
using these adapters: -
1. Screw the appropriate gauge adapter
fully on to the instrument to be tested.
2. Screw assembly down COUNTER-CLOCKWISE
on to test port.
Note:
Hand-tight is sufficient; ensure that the bottom
face contacts the O-ring on the test port.
FIGURE 2-2A
MAKING PRESSURE CONNECTIONS
2-2
CONNECTIONS