Section 7-7 - Troubleshooting Trees, Instructions and Tech Tips
7-7-23
Electromagnetic Interference (EMI) Troubleshooting
7-7-23-1
Prevention/abatement
For EMI rules and details, see "EMI Limitations" on page 2-4.
7-7-23-2
Different power outlet
Connect the LOGIQ™ S8 to another power outlet and check to see if the noise changes or disappears.
NOTE:
Image artifacts can occur if, at any time within the facility, the ground from the main facility's incoming
power source to the LOGIQ™ S8 is only a conduit.
See "Minimal floor plan suggestions" on page 2-9.
7-7-23-3
Different system
Try another LOGIQ™ S8 at the same location and look for the same noise. If the noise is present on
the new system too, the noise is most likely from an external source/equipment.
7-7-23-4
Different location
Move the LOGIQ™ S8 to another location and check to see if the noise changes or disappears. This
may help to locate an external noise source.
Try to move the LOGIQ™ S8 to:
•
another location inside the room
•
another room
•
another floor
7-7-23-5
Disconnect external cables
Disconnect all external cables (network, all unused probes, ECG leads, etc.) and check to see if the
noise disappears.
Содержание 5418099
Страница 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO ...
Страница 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x ...
Страница 16: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL xiv This page was intentionally left blank ...
Страница 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16 ...
Страница 176: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 98 Section 3 16 Paperwork ...
Страница 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4 ...
Страница 417: ......