Section 7-7 - Troubleshooting Trees, Instructions and Tech Tips
7 - 71
7-7-5
Trackball - Impaired sensitivity
Figure 7-64 Trackball - Impaired sensitivity
Trackball has low sensitivity.
Set up the appropriate trackball response time.
(via [Speed] and [Acceleration] in the Utility - System - General
- Trackball. Refer to
Is the trackball issue resolved?
Is the trackball issue resolved?
Clean the trackball.
Refer to
Replace the trackball.
Return to normal scanning.
Return to normal scanning.
No
No
No
Yes
Yes