Sensor Information
GE Sensing produces a variety of sensors compatible with the
OptiSonde, ranging from one to two stages of thermoelectric cooling.
A comparison chart listing specifications of each sensor appears in
Appendix D. The following sections provide information on
installing the following GE Sensing dew point sensors:
•
Model 1111H — Single-stage sensor with 1” M-NDT
•
Model 1111H-GE — Dew point single-stage sensor with flange
•
Model 1211H — Two-stage sensor; for high pressure and temp.
•
Model D-2 — Two-stage sensor
When selecting a location for installing a sensor, consider the
following criteria:
•
Locate the sensor as close as is practical to the source of the gas to
be measured, to keep the sampling lines as short as possible. This
minimizes the system response time and reduces the error rate at
low frost points due to sample line outgassing.
Caution!
The OptiSonde electronics and sensors are for General
Purpose Locations only. They are not for use in hazardous
(classified) locations. Sample air/gas shall be non-
combustible only.
•
Choose a sensor location that provides access to the dewpoint
sample cavity cover, to facilitate periodic mirror cleaning.
Caution!
Never place the sensor in a location where temperatures
rise above the maximum rated temperature for the device.
See Appendix D for complete sensor specifications.
Содержание 1111H
Страница 1: ...GE Sensing OptiSonde General Eastern Chilled Mirror Hygrometer User s Manual ...
Страница 9: ...Chapter 1 ...
Страница 10: ...Features and Capabilities Introduction 1 1 Electronics Enclosure 1 2 The System 1 3 Sensors 1 4 ...
Страница 15: ...Chapter 2 ...
Страница 33: ...Chapter 3 ...
Страница 43: ...Chapter 4 ...
Страница 44: ...Programming the OptiSonde Introduction 4 1 Programming Technique 4 2 Programmable Functions 4 3 ...
Страница 57: ...Chapter 5 ...
Страница 68: ...Appendix A ...
Страница 73: ...Appendix B ...
Страница 74: ...Humidity Equations and Conversion Chart Introduction B 1 Vapor Pressure B 1 Humidity B 2 ...
Страница 77: ...Humidity Equations and Conversion Chart B 3 October 2007 Figure B 1 Graphical Humidity Conversion Chart ...
Страница 78: ...Appendix C ...
Страница 79: ...Configuring the Serial Interface Wiring to a Personal Computer C 1 ...
Страница 82: ...Appendix D ...
Страница 83: ...Chilled Mirror Sensors Introduction D 1 Depression D 1 Measurement Range D 2 Comparing OptiSonde Sensors D 3 ...
Страница 87: ...Appendix E ...
Страница 88: ...Theory of Operation and Glossary Theory of Operation E 1 The PACER Cycle E 4 Glossary E 5 ...