Functional and electrical safety checks:
Functional Checkout procedures
2000966-542D
Dash 3000/4000/5000
8-27
Cardiac output tests (option)
Equipment
A multiparameter patient simulator
Cardiac output cable adapter
Cardiac output simulator
Connections
1. Connect the cardiac output cable adaptor to the
TEMP/CO
connector
of the patient monitor.
2. Connect the cardiac output adapter to the cardiac output simulator.
3. Verify a cardiac output parameter window appears on the patient
monitor.
Cardiac output test procedures
1. On the patient monitor, select CO Parameter Box.
2. Set up the patient monitor:
Set
AUTO MODE:
to
ON
.
Set
INJECT TEMP:
to
BATH
.
Set
SIZE:
to
7
.
Set
INJECT VOL:
to
10CC
.
Set
COMPUTATIONAL CONSTANT:
to
0.540
.
3. Set cardiac output cable adapter or simulator to
0°
.
4. Turn the simulator on. Select cardiac output menu corresponding to
the injectate temperature set in step 3 at 5 liters/minute
5. When the patient monitor screen displays the message,
INJECT
WHEN READY
, press appropriate button on simulator to inject.
6. When computing is complete, the CO reading should be
approximately 37°C at 5 liters of blood per minute ± 5%.
Invasive blood pressure tests (option)
Equipment
A multiparameter patient simulator
Dual BP cable (2005772-001 or equivalent)
The invasive blood pressure (BP) tests provide a method of verification
for all BP connectors (
BP1
,
BP2
,
BP1/3
and
BP2/4
) of a patient monitor
equipped with this optional function.
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......