Field replaceable units:
Replace display assembly parts
2000966-542D
Dash 3000/4000/5000
7-27
Open display assembly
Follow the steps below for the patient monitor you are servicing.
Dash 3000 patient monitor
with display shield flex circuit
Dash 4000 patient monitor
Dash 5000 patient monitor
Separate the patient monitor bezel from
the LCD display.
1. Remove and retain the four screws
holding the display shield to the LCD
display and two screws holding the
display shield to the patient monitor
bezel.
2. Remove the display assembly from the
patient monitor bezel.
3. Pull the LCD display isolator out of the
way. Disconnect the LCD display cable.
501A
4. Disconnect the cable from the backlight
inverter PCB.
502A
1. Remove and retain the two screws from
the display flex cable. Pull up on the
latch tabs to remove the flex cable from
the display.
2. Remove and retain the 10 screws as
shown.
3. Separate the transition bezel from the
front bezel. It is not necessary to remove
the metal backplate.
4. Set aside the transition bezel.
NOTE
The display that is pictured above is a
Sharp.
1. Remove and retain the two screws from
the display flex cable. Pull up on the
latch tabs to remove the flex cable from
the display.
2. Remove and retain the 14 screws as
shown.
3. Separate the transition bezel from the
front bezel. It is not necessary to remove
the metal backplate.
4. Set aside the transition bezel.
2 smaller screws
4 larger screws
500A
LCD display cable
cable
879A
2 screws
10 screws
899A
910A
860A
2 screws
14 screws
900A
884A
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......