Measurement Formulas
2
-
22
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
Direction 5116269-100 Rev. 2
RtOvFo[ml]
Right Ovary Follicles
One distance
D1[cm]
3
x
π
/6
Two distances
D1[cm]
2
x D2[cm] x
π
/6
: (D1 <D2)
D1[cm] x D2[cm]
2
x
π
/6
: (D2 <D1)
Three distances
D1[cm] x D2[cm] x D3[cm] x
π
/6
a.V
diastole
= V
min
or V
end-diastole
(depends on preset selection)
b.To calculate LtOvFo or RtOvFo with one (or two distances), press the
Clear
key after the first (or
second distance) measurement(s).
Table 2-9: GYN Calculation Formulas
Calc
Mnemonic
Calc Name
Input
Measurements
Formula