Quick Start 18
LOGIQ
t
400 Quick Start Guide
2237879–100 Rev. 0
Color Flow Mode (CFM) Controls (cont’d)
Packet Size—Controls the number of samples gathered for a single color
flow vector.
Spatial Filter—Smooths color information so that it is less grainy. Averages
the color information out over time.
W. E. Cancel—Eliminates the low velocity echoes caused by the motion of
vessel walls.
Color Tag—Allows for the assignment of a single color to a range of
velocities.
Tag Positn—Allows for the movement of the specified color tag range
throughout the gray scale displayed.
ACE—Reduces “color artifact noise”.
Noise Blanker—Reduces the random noise generated due to the increase in
Color Gain.
Persistence—Retains the largest pixel color value until a larger value is
detected or preset time has elapsed.
Tag
Positn
Color
Tag
Packet
Size
CFM
3/4
W.E.
Cancel
Spatial
Filter
SMALL
OFF
Persist
ence
4/4
Noise
Blanker
ACE
CFM
OFF
OFF
OFF