5-8
Model 340 Telemetry System
Revision A
2006920-001
Calibration: Receiver Calibration
Mode Controls
1.
Confirm the low-pass filter –3 dB cutoff point of 90 ± 6Hz with an
input of 1.0 Vp-p at TP45 and the output at TP21. Set R61 for a gain
of 2.3.
2.
Adjust R177 for a frequency of 8020 ± 20Hz at TP32.
3.
Confirm a 10 ± 0.1 Vp-p, 4010 ± 20 Hz square wave at TP33.
4.
Confirm the center frequency and the –3 dB points of the switched
capacitor bandpass filters for each IC listed in
.
5.
With no signal applied, adjust the tone decoders for the correct center
frequency indicated in
.
dB
UNITS
AC
VOLTS
DC
VOLTS
888 888 888
Hz Generator
888 888 888
Hz Counter
Table 5-2. Band-Pass Filter Center Frequencies and –3 dB Points
IC Reference
Center Frequency
–3 dB Point
U17-6
80 Hz
75 Hz, 85 Hz
U18-6
40 Hz
35 Hz, 45 Hz
U23-6
20 Hz
15 Hz, 25 Hz
U27-6
10 Hz
5 Hz, 15 Hz
Table 5-3. Center Frequencies of Tone Decoders
Test Point
Center Frequency
Adjustment
TP6
10 Hz
R95
TP8
20 Hz
R99
TP9
40 Hz
R110
RP11
80 Hz
R149
Содержание Corometrics 340
Страница 2: ......
Страница 8: ...iv Model 340 Telemetry System Revision A 2006920 001 For your notes ...
Страница 10: ...For your notes ...
Страница 12: ...For your notes ...
Страница 30: ...2 4 Model 340 Telemetry System Revision A 2006920 001 For your notes ...
Страница 66: ...5 14 Model 340 Telemetry System Revision A 2006920 001 For your notes ...
Страница 70: ...6 4 Model 340 Telemetry System Revision A 2006920 001 For your notes ...
Страница 74: ...7 4 Model 340 Telemetry System Revision A 2006920 001 For your notes ...
Страница 80: ...9 2 Model 340 Telemetry System Revision A 2006920 001 For your notes ...
Страница 81: ......
Страница 82: ...MANUAL P N 2006920 001 REV A Ê2006920 001ÅŠ ...