System Setup
Critical Values Notification Setup (cont'd.)
Field
Description
Ischemia detection
Ischemia detection
enables the notification for ECGs in which
ischemia is detected. Select
Yes
to enable ischemia notifications.
AV Block detection
AV Block detection
enables the notification for ECGs in which an AV
block is detected. Select
Yes
to enable AV Block notifications.
Arrhythmia
detection
Arrhythmia detection enables the notification for ECGs in which
an arrhythmia is detected. Select
Yes
to enable arrhythmia
notifications.
NOTE:
For more information on the critical values for STEMI detection, ischemia detection,
AV block detection, or arrhythmia detection, refer to the
12SL Physician's Guide
.
Saving Critical Values to SD Card
Use the following procedure to save your critical value notification settings to an
external SD card. You can use the card to restore the notification settings in the case
of a system failure. You can also use the card to quickly configure other systems.
1.
Configure your
Critical Values Notifications
.
Refer to
“Notifications Setup” on page 137
for details.
2.
When you are done configuring your settings, select
Return
at the bottom of
the window.
This saves your settings to local memory and returns you to the
Critical Values
Setup
menu.
3.
Insert the SD card if necessary.
4.
Select
Save to SD Card
and press
Return
.
The settings are saved to the card.
5.
Eject the card and save it in a safe location or use it to configure other devices.
Restoring Critical Values from SD Card
Use the following procedure to restore your critical value notification settings from an
external SD card. You can use this procedure to restore the notification settings in the
case of a system failure. You can also use it to quickly configure other systems.
1.
Insert the SD card with your critical value notification settings.
Refer to
“Saving Critical Values to SD Card” on page 138
for details and backing
up your settings to an SD card.
2.
From the
Main Menu
select
System Setup > ECGs > Critical Values Setup >
Restore from SD Card
.
The settings on the SD card overwrite any settings on the device.
3.
Eject the card and save it in a safe location or use it to configure other devices.
138
MAC
®
5500/MAC
®
5500 HD
2046275-016B
Содержание MAC 5500
Страница 54: ...Preparing the Patient 54 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 62: ...Entering Patient Information 62 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 80: ...Editing Protocols 80 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 98: ...Deleting ECGs and Orders 98 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 167: ...Maintenance 2046275 016B MAC 5500 MAC 5500 HD 167 ...
Страница 168: ...Maintenance 168 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 178: ...Report Formats 178 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 187: ......