Recording a Resting ECG
Continue with step
5.
Press
Cancel
to stop acquiring the ECG, and perform one of the following:
• Press
Yes
to stop acquiring the ECG data and return to the
Resting ECG
window.
• Press
No
to remain acquiring the ECG data, and continue with step
6.
In the
Full Disclosure Report
window, you can perform the following operations:
Full Disclosure Report Window Field
Action
Export to SD Card
Press to export the ECG report to the SD
card.
Export to FTP Server
Press to export the ECG report to the
configured FTP Server.
See
“Communication Setup” on page 140
for detailed instruction on configuring the
FTP server.
Export to Shared Folder
Press to export the ECG report to the
configured Shared Folder.
See
“Communication Setup” on page 140
for detailed instruction on configuring the
Shared Folder.
Press to print the ECG report to the
thermal printer.
Report Setup
Press to setup the full disclosure report.
See
“Full Disclosure Report Setup” on
page 79
for detailed instructions.
Return
Press and then the following message
displays:
The current full disclosure record will be
deleted. Do you want to continue?
• Press
Yes
to return to the Resting ECG
window.
• Press
No
to remain in the current full
disclosure report window.
78
MAC™ 2000
2053535-002K
Содержание MAC 2000
Страница 30: ...Introduction 30 MAC 2000 2053535 002K ...
Страница 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Страница 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Страница 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Страница 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Страница 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Страница 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Страница 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Страница 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Страница 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Страница 224: ...224 MAC 2000 2053535 002K ...
Страница 225: ......