Maintenance and checkout
129
2062973-002
10 Maintenance and checkout
This chapter specifies the checkout procedure and the maintenance activities to be performed
to the patient monitor after corrective maintenance and during annual planned maintenance.
WARNING
Only perform maintenance procedures specifically described in the manual.
WARNING
Planned maintenance should be carried out annually. Failure to implement
the recommended maintenance schedule may cause equipment failure and
possible health hazards.
NOTE: The manufacturer does not, in any manner, assume the responsibility for performing the
recommended maintenance schedule, unless an Equipment Maintenance Agreement exists.
The sole responsibility rests with the individuals, hospitals, or institutions utilizing the device.
NOTE: Refer to Module Frames and Modules Technical Manual for corrective and planned
maintenance checkout procedures of the parameter modules.
NOTE: Refer to the PDM section in the Module Frames and Modules Technical Manual for the
battery maintenance procedure of the PDM battery.
Corrective maintenance
Service personnel shall perform the following checkout procedure steps after any corrective
maintenance, before taking the monitor back into clinical use:
Performed service activity
Required checkout procedure
Visual
inspections
(section 10.1)
Electrical
safety test
(section 10.2)
Functional check
(section 10.3)
After detaching, replacing or
upgrading:
- Recorder Unit FRU / Upgrade
- E-module Board FRU / Upgrade
- Module Frame Assembly FRU
Yes
Yes
- 10.3.1 Start-up
- 10.3.3 PSM / PDM
- 10.3.4 E-module identification
- 10.3.13 Recorder
After detaching or replacing:
- PDM Docking Mechanism FRU
- Module Frame Cover Unit Set FRU
Yes
No
- 10.3.1 Start-up
- 10.3.3 PSM / PDM
After detaching, replacing or
upgrading:
- Standard Interface Board FRU
- Advanced Interface Board FRU /
Upgrade
Yes
Yes
- 10.3.1 Start-up
- 10.3.6 Mouse
- 10.3.7 Alphanumeric
After replacing or upgrading:
- Wireless LAN FRU / Upgrade
Yes
No
- 10.3.1 Start-up
- 10.3.10 Wireless LAN
After replacing:
- Mains Fuses FRU
- Battery FRU
Yes
No
Содержание CARESCAPE Monitor B650
Страница 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Страница 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Страница 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Страница 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Страница 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Страница 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Страница 177: ...Troubleshooting 169 2062973 002 ...
Страница 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Страница 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Страница 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Страница 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Страница 239: ......