5 Methods
5.6 Cycle types
92
Biacore T100 Software Handbook BR-1006-48 Edition AE
The
High performance
and
Low sample consumption
options support a
Mix
function for mixing sample with a defined solution in the autosampler before
injection. Check the
Mix
option and enter a mix solution and mixing fraction to
use this function. Entering a fraction of e.g. 20% will mix one part of mixing
solution with four parts of sample. The sample and mixing solution are taken
from respective positions in the autosampler and mixed in a third position. The
option
Stabilization period after mix
allows you to specify a wait period
between the mixing operation and injection of the mixed solution.
Mix
is not
supported for the
Single cycle kinetics
option.
Solvent correction command
This command injects a 30-second pulse of solvent correction solution over the
surface at a flow rate of 30 µl/min. A solvent correction cycle should contain 4-8
Solvent correction
commands for the different solvent concentrations used to
construct the correction curve (see Section 6.7).
Solvent correction
commands
will be correctly evaluated only when they are used in cycle types that are
connected to assay steps with purpose
Solvent correction
.
5.6.2
Variables
Parameters for many of the commands in a cycle definition may be set as
variables. Values for variables are entered in either the
Variable Settings
in the
method or the
Setup Run
step when the method is run (Sections 5.7 and 5.9.2),
and determine the number of cycles that will be performed in the run. Variables
fall into two broad classes:
•
Method variables
such as sample name or contact time control the way
the cycle will be performed. Parameter values that are not set as variables
are defined in the main command panel.
•
Evaluation variables
such as concentration or analyte molecular weight
are used in evaluation of the data. Some evaluation variables are required
for correct functioning of application-specific evaluation procedures (for
example, kinetic evaluation requires a variable called
Conc
which holds
the analyte concentration). These are selected from a predefined list.
Other evaluation variables may be freely defined by the user, to hold
Single cycle kinetics
Injects a series of sample concentrations in the same
cycle, intended for single-cycle kinetics analysis (see
Section B.9). The samples are injected in direct
sequence, separated only by the time required to
prepare the next injection. A dissociation time is
included after the last sample injection. With this option
checked, an evaluation variable (see Section 5.6.2) will
be set up for each of the specified number of
concentrations.
Содержание Biacore T100
Страница 1: ...GE Healthcare Biacore T100 Software Handbook ...
Страница 2: ......
Страница 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Страница 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 24: ...2 Control Software general features 2 4 File storage 24 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 76: ...4 Application wizards 4 6 Control experiment wizards 76 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Страница 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 138: ...7 Data presentation tools 7 4 Report point table 138 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Страница 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 214: ...Appendix A Data import and export A 2 Importing data 214 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 237: ......