Biacore T100 Software Handbook BR-1006-48 Edition AE
71
Application wizards 4
changed does not interfere with the first measurements. Start-up cycles will be
run at each temperature.
Check
Run solvent correction
to include solvent correction cycles in the run.
Solvent correction adjusts response values for the effects of varying bulk
refractive index contribution between samples, and may improve the quality of
the results for analyses using small organic analytes that require dimethyl
sulfoxide (DMSO) in the sample buffer to maintain solubility. The principles and
application of solvent correction are described in Section 6.7.
Set the required number of injections per cycle and the frequency of solvent
correction cycles. The default settings use 8 injections per cycle and repeat the
solvent correction cycle every 15 sample cycles. Separate solvent correction
cycles will be run at each temperature.
Enter the temperatures at which the measurements are to be performed. For
most purposes, 5-7 temperatures will be adequate: fewer points make the
determination of thermodynamic parameters uncertain, while more points
increase the run time. Distribute the points evenly over the widest temperature
interval that the ligand and analyte tolerate. Start from the lowest temperature
to minimize the time needed for temperature equilibration between
measurements (increasing the analysis temperature takes less time than
decreasing it). The system will wait for a stable temperature between each
determination. As an additional control, you may want to include a replicate of
the first temperature at the end of the run.
Note:
For robust determination of standard thermodynamic parameters (
Δ
G°,
Δ
H° and
Δ
S°), arrange the analysis temperatures so that 25°C is in the
middle of the range (e.g. 10–40°). See Section 10.1 for further details.
Specify the sample compartment temperature, or check
Vary with analysis
temperature
to change the sample compartment temperature automatically
when the analysis temperature is changed.
Step 3. Injection parameters
Specify the injection parameters for each injection in the cycle (see
Section 4.2.3). The sample injection for thermodynamics measurement has an
additional setting for dissociation time. This is the time for which dissociation will
be monitored after the end of the injection without disturbances from flow
system washing procedures.
Step 4. Samples
Enter the details of the samples. This dialog box is equivalent to the
concentration series dialog for kinetics determination (see Section 4.5.3, step 4).
Separate rack positions will be created for each sample at each temperature.
Содержание Biacore T100
Страница 1: ...GE Healthcare Biacore T100 Software Handbook ...
Страница 2: ......
Страница 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Страница 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 24: ...2 Control Software general features 2 4 File storage 24 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 76: ...4 Application wizards 4 6 Control experiment wizards 76 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Страница 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 138: ...7 Data presentation tools 7 4 Report point table 138 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Страница 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 214: ...Appendix A Data import and export A 2 Importing data 214 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 237: ......