Biacore T100 Software Handbook BR-1006-48 Edition AE
35
Application wizards 4
the chip has been stored with immobilized molecule and special treatment is
required to wash out the storage medium.
If you run several assays after each other without undocking the chip between
assays, conditioning is generally only required for the first assay.
Start-up cycles
Start-up cycles are identical to analysis cycles except that the sample is
replaced by a dummy sample (often buffer). The response from a newly
prepared or newly docked sensor chip often shows some instability during the
first few cycles, and start-up cycles allow the response to stabilize before the
first analysis cycle is performed. Three start-up cycles are generally
recommended for most assay purposes, to ensure a stable response in the
analysis. Start-up cycles are treated separately from analysis cycles in the
evaluation software.
Start-up cycles are run at the beginning of the experiment (after the
conditioning cycle), and also directly after buffer change (in the
Buffer Scouting
wizard) and temperature change (in the
Thermodynamics
wizard).
4.2.3
Injection parameters
The
Injection parameters
dialog specifies details of the injections selected in
the
Injection sequence
.
Details of this dialog box may vary according to the injections selected and the
particular wizard. Some features may be generalized:
Содержание Biacore T100
Страница 1: ...GE Healthcare Biacore T100 Software Handbook ...
Страница 2: ......
Страница 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Страница 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 24: ...2 Control Software general features 2 4 File storage 24 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 76: ...4 Application wizards 4 6 Control experiment wizards 76 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Страница 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 138: ...7 Data presentation tools 7 4 Report point table 138 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Страница 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 214: ...Appendix A Data import and export A 2 Importing data 214 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 237: ......