4 Application wizards
4.5 Assay wizards
68
Biacore T100 Software Handbook BR-1006-48 Edition AE
Step 4. Samples
Enter the details of the samples for kinetic or affinity determination. For each
analyte, a zero concentration sample and at least four non-zero concentrations,
one of which is run in duplicate, are strongly recommended. Concentrations are
entered in the left-hand
Concentration
column, either in molar or weight-based
units. Choose the concentration units from the pull-down list in the column
header. If a weight-based concentration unit is chosen (e.g. µg/ml) a molecular
weight must also be specified. When a molecular weight is entered, the right-
hand
Concentration
column displays the conversion from molar to weight-
based or
vice versa
. Samples with the same sample name may not be given
different molecular weights.
The samples may be analyzed either in the order entered in the table or sorted
in increasing concentration. The order displayed in the sample table is not
affected by the choice of run order.
If you enter samples with different names, they will be handled as separate
concentration series regardless of the order in which they are entered. The
samples will be run as separate concentration series even if the order is mixed
in this dialog: thus samples entered in the order A, B, A, A, B, A, B, B… will be run
in the order A, A, A, A, B, B, B, B… The
Run order
setting applies within each
concentration series.
Содержание Biacore T100
Страница 1: ...GE Healthcare Biacore T100 Software Handbook ...
Страница 2: ......
Страница 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Страница 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 24: ...2 Control Software general features 2 4 File storage 24 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 76: ...4 Application wizards 4 6 Control experiment wizards 76 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Страница 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 138: ...7 Data presentation tools 7 4 Report point table 138 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Страница 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 214: ...Appendix A Data import and export A 2 Importing data 214 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 237: ......