Cleaning and care
9-1
9 Cleaning and care
Safety precautions
Warnings
•
Disconnect equipment from power line before cleaning
•
If liquid has accidentally entered the system or its parts, disconnect the power cord from
the power supply and have the equipment serviced by authorized service personnel.
•
Regular preventive maintenance should be carried out annually.
•
Do not use unspecified cleaners, materials or chemicals as they may damage device
surfaces, labels, or cause equipment failures.
•
Do not pour or spray any liquid directly on cables or leadwires or permit fluid to seep into
connections or openings.
•
Never use conductive solutions, solutions that contain wax, or wax compounds to clean
devices, cables or leadwires.
•
Do not immerse any part of the device in liquids or allow liquid to enter the interior.
•
The decision to sterilize must be made per your institution’s requirements with an
awareness of the effect on the integrity of the cable or leadwire.
•
Clean the surface of the probe before and after each patient use.
•
CABLE/SENSOR AFTER CARE
−
Do not immerse sensors or patient cables in water, solvents or cleaning solutions.
−
Do not reuse sensors intended for single patient use.
−
Do not sterilize sensors or patient cables by irradiation, steam, or ethylene oxide.
−
Clean the surface of the probe before and after each patient use.
−
Allow sensor and cable to dry completely after cleaning. Moisture and dirt on the
connector can affect the measurement accuracy.
−
If a probe is damaged in any way, discontinue use immediately.
−
Inaccurate SpO
2
data can result if a sensor is past its useful life.
−
A damaged sensor or a sensor soaked in liquid may cause burns during
electrosurgery.
•
Since calibration gas contains anesthetic agents, always ensure sufficient ventilation of
the room during calibration.
•
Inaccurate readings due to:
−
use of unapproved accessories
−
reuse of single-use accessories
−
affect on pressure strong scavenging suction
•
The user may only perform maintenance procedures specifically described in this
manual.
•
On the gas module indicates that airway gases should be calibrated every six months in
normal use and every two months in continuous use.
Содержание B20
Страница 2: ......
Страница 16: ...B40 B20 Patient Monitor xii ...
Страница 17: ...1 Introduction ...
Страница 18: ......
Страница 27: ...2 System description ...
Страница 28: ......
Страница 58: ...B40 B20 Patient Monitor 2 30 ...
Страница 59: ...3 Installation ...
Страница 60: ......
Страница 69: ...4 Monitoring basic ...
Страница 70: ......
Страница 79: ...5 Alarms ...
Страница 80: ......
Страница 102: ...B40 B20 Patient Monitor 5 22 ...
Страница 103: ...6 Monitor setup ...
Страница 104: ......
Страница 123: ...7 Trends ...
Страница 124: ......
Страница 140: ...B40 B20 Patient Monitor 7 16 ...
Страница 141: ...8 Print and record ...
Страница 142: ......
Страница 153: ...9 Cleaning and care ...
Страница 154: ......
Страница 163: ...10 Troubleshooting ...
Страница 164: ......
Страница 181: ...11 ECG ...
Страница 182: ......
Страница 205: ...12 Impedance respiration ...
Страница 206: ......
Страница 215: ...13 Pulse oximetry ...
Страница 216: ......
Страница 232: ...B40 B20 Patient Monitor 13 16 ...
Страница 233: ...14 Non invasive blood pressure ...
Страница 234: ......
Страница 248: ...B40 B20 Patient Monitor 14 14 ...
Страница 249: ...15 Invasive blood pressure ...
Страница 250: ......
Страница 261: ...16 Temperature ...
Страница 262: ......
Страница 267: ...17 Airway gas ...
Страница 268: ......
Страница 293: ...18 Entropy ...
Страница 294: ......
Страница 316: ...Abbreviations A 12 ...
Страница 318: ...B40 B20 Patient Monitor B 2 ...
Страница 319: ......