ZX3030B manual
- 11 -
DC output operation
Figure 8:DC operating window
Voltage and current range setting:
Press the corresponding [< <] key and highlight the
voltage range 10.0V, change the range by the up and down key, and press the [confirm] key to
execute. Press the [< <] key once to enter the current path setting. And setting is the same
way.
Voltage range shortcut key definition:
Select range directly through numeric
keys:1:75mV, 2:150mV, 3:300mV, 4:3V, 5:6V, 6:10V, 7:75V, 8:150V, 9:300V, 0:500V
Current path shortcut key definition:
Select range directly through numeric keys:1:30uA,
2:300uA, 3:3mA, 4:30mA, 5:300mA, 6:1A, 7:5A, 8:20A
Voltage and current amplitude adjustment:
According to the corresponding key[< <] , the
voltage amplitude 00.000% is highlighted. The right and left keys move the cursor to change
the setting fineness, and the up and down key to change the amplitude, and the output
amplitude is set according to [output] key. Enter the current range setting by pressing the [< <]
key once
Voltage and current amplitude shortcut key definition:
1:10%, 2:20%, 3:30%, 4:40%,
5:50%, 6:60%, 7:70%, 8:80%, 9:90%, 0:100%