background image

 

CLEANING & MAINTENANCE 

Fig 3. GASMAX II + GDS-IR  

Apply power to the GASMAX II + GDS-IR and allow the 
instrument to stabilize for at least one (1) hour. 

 

IMPORTANT 

- Be sure there are no hydrocarbon gases 

present during the IR Zero procedure. If necessary, apply 
a flow of nitrogen or known ‘zero air’ to the Cal Gas Inlet 
(See Fig 3.). 

Step 1: Setting IR Zero on the GDS-IR 

Open the GASMAX II enclosure cover and loosen the two 
thumbscrews that hold the display board in place. While 
holding  the  display  in  one  hand,  PRESS  and  HOLD  the 

IR  Zero 

switch for 12 to 15 seconds, but not more than 

20 seconds. If the GASMAX II Marker option is turned on, 
the  GASMAX  II  display  will  momentarily  show  the 
message  “IR  ZERO”  when  the  GDS-IR  output  drops  to 
2.2mA to indicate a successful zero setting. Replace the 
display assembly and tighten the thumbscrews.  

Step 2: Calibrating the GASMAX to the GDS-IR 

Hold a magnetic wand over the DN / CAL button on the 
lower left side of the GASMAX II display. When the CAL? 
Message appears, confirm by holding the wand over the 
EDIT  key.  Follow  the  instructions  that  appear  on  the 
GASMAX  II  display  to  fully  calibrate  the  GASMAX  + 
GDS-IR by applying zero and span gas to the sensor.  

INITIAL STARTUP & CALIBRATION 

Periodically open and inspect the of the hydrophobic filter 
by removing the Outer Barrel Assembly.  
The  hydrophobic  filter  is  a  Teflon-coated  stainless  steel 
mesh  that  keeps  moisture  and  particulates  out  of  the 
optical  path.    A  setscrew  holds  the  filter  to  the  GDS-IR 
housing.  Compressed air can be used to clean the filter 
of dust and moisture.  
To  inspect  or  clean  the  optical  path,  remove  the 
hydrophobic  filter  by  unscrewing  the  setscrew  at  the 
bottom of the filter assembly. The filter will then slide off, 
revealing the optical waveguide assembly.  
The waveguide and waveguide collar can be removed by 
inserting rigid instruments (Allen wrenches) into the holes 
of the waveguide and collar.  Turning the waveguide and 
collar  in  opposite  directions  will  loosen  the  waveguide 
allowing  the  collar  to  be  screwed  down  onto  the 
waveguide  until  both  can  be  removed  from  the  GDS-IR 
housing. Clean the two sapphire windows with a lint-free 
cloth and reassemble the unit.  
Always perform an IR zero procedure as described above 
if the unit is physically moved or disassembled.  

S2 

Ribbon to 

P

W

R

 

O

U

T

 4

-2

“E

C

H

E

M

” 

O

U

T

 4

-2

“L

E

L”

 

G

N

D

 

R

E

F

 

C

O

M

 

CATBEA

S

C7 

C8 

C1 

C2 

I/O Power 

Supply 

TB1 

TB2 

IR 
Zero 
Switch 

Cal 
Gas 
Inlet 

GDS-IR 

Infrared Gas 

Sensor 

Outer 

Barrel 

Assembly 

GASMAX II 

Gas Monitor 

Отзывы: