
GASMAX CX Operation & Maintenance Manual, Revision 1.0
Page 32
8
TROUBLESHOOTING
TOXIC
SENSOR
INDICATES
FAULT
OR
OVERRANGE
Certain
toxic
sensors
indicate
off
‐
scale
low
or
high
at
power
up
and
quickly
drift
towards
zero.
This
is
normal
behavior.
Toxic
sensors
showing
constant
FAULT:
If
local,
remove
sensor
and
examine
for
moisture
or
discoloration.
Replace
sensor
if
wet
or
discolored.
If
remote,
check
sensor
cable
and
junction
box
for
moisture
or
standing
water.
Remove
sensor
and
examine
for
moisture
or
discoloration.
FAULT
indication
generally
indicates
sensor
useful
life
is
exhausted.
Toxic
sensors
left
unpowered
for
more
than
3
months
are
subject
to
accelerated
degradation
and
may
demonstrate
a
permanent
loss
of
sensitivity.
TOXIC
SENSOR
WILL
NOT
CALIBRATE
Sensor
reading
during
zero
calibration
exceeds
upper
limit
of
zero
–
sensor
is
defective
and
should
be
replaced.
Sensor
reading
during
span
calibration
too
low
–
sensor
may
be
defective.
However,
it
may
be
possible
to
temporarily
continue
operation
by
increasing
PREAMP
GAIN.
BRIDGE
SENSOR
INDICATES
FAULT
OR
OVERRANGE
Catalytic
bead
combustible
sensors
generally
indicate
off
‐
scale
high
at
power
up
and
quickly
drift
towards
zero
as
they
reach
operating
temperature.
This
is
normal
behavior.
Combustibles
sensors
showing
constant
OVERRANGE
may
have
defective
bead.
If
remote,
check
wiring
for
correct
excitation
voltage
at
sensor.
Replace
sensor.
BRIDGE
SENSOR
WILL
NOT
CALIBRATE
Sensor
reading
during
zero
calibration
exceeds
limits
–
readjust
SENSOR
BALANCE
to
reset
zero
if
possible.
If
not,
sensor
is
defective
and
should
be
replaced.
Sensor
reading
during
span
calibration
too
low
–
sensor
may
be
defective.
However,
it
may
be
possible
to
temporarily
continue
operation
by
increasing
PREAMP
GAIN.
See
Section
8.2
for
more
details.