GDS Corp GASMAX II Скачать руководство пользователя страница 2

GDS Corp. 

2513 Hwy 646 

Santa Fe, Texas 77510 

(409) 927-2980  (409) 927-4180 fax    

www.gdscorp.com   [email protected] 

SPARE PARTS & ACCESSORIES 

10-9060 

Replacement PID Sensor (0-2000ppm) 

10-9061  

Replacement PID Sensor (0-20ppm) 

10-906XR 

Sensor Refurbishment for PID 

10-0198 

Sensor splash guard 

10-0203 

Sensor calibration cup 

10-0205 

Flow cell for process monitoring 

10-0187 

Sensor removal tool 

1000-0078 Large 

black 

magnet 

10-0232 Replacement 

display 

10-0233 

Replacement I/O power supply  

10-0234 

Replacement Alarm / MODBUS  

10-0250 

Replacement isolated 4-20mA 

If a 

local 

PID smart sensor is installed, the GASMAX II 

will automatically identify the sensor type during 
power-up and load the configuration parameters 
stored in the sensor.  

V. LOCAL SENSOR SETUP 

Calibration should be performed at least once every 
three months for PID sensors. In some cases, more 
frequent calibration may be required due to 
environmental considerations.  
Attach a calibration cup to the sensor head and apply 
a steady flow of “zero air” to the instrument for one 
minute. Hold the magnetic wand over the “DN / CAL” 
switch on the lower left side of the display until the 
GASMAX enters calibration mode. Follow the on-
screen instructions to calibrate the GASMAX II unit.  

If a remote PID sensor is installed, sensor information 
must be entered manually.  

ENABLE THE LEL/CURRENT CHANNEL 

 

Press and hold the UP key for 5 seconds until the 
“Enter Key Sequence” menu appears. Press the UP 
key four times in sequence to access the XMITTER 
CONFIG menu. Make sure the LEL/CURRENT 
channel is ON. If an optional toxic sensor is also 
installed (two-channel mode), make sure the O2/
TOXIC channel is also ON. Exit the XMITTER 
CONFIG menu by pressing the NEXT key.  

INITIALIZE THE GASMAX SOFTWARE 

Press the EDIT key to display the Main Menu. Select 
the Sensor Information menu entry and press EDIT. 
Select INSTALL SENSOR and press EDIT to confirm. 
Select Cat-Bead Input Type. This step deletes any 
existing historical calibration data and resets sensor 
life to 100%. Press NEXT to exit the menu.  

SET SENSOR VOLTAGE 

Press  the EDIT key to display the Main Menu. Scroll 
down to the TECHNICIAN ONLY entry and press 
EDIT again. Press the UP key four times in sequence 
to enter the TECHNICAN ONLY menu.  
Select SET SENSOR VOLTAGE and adjust the 
catalytic bead excitation voltage to 5.0V as measured 
across the REMOTE SENSOR JUNCTION BOX 
TERMINAL STRIP “A” and “R” terminals.  

IMPORTANT: Do not exceed 5.5VDC across the 
PID sensor at any time.  Excess voltage can 
damage the sensor and void the warranty,  

VI. REMOTE SENSOR SETUP 

Press NEXT to exit the TECHNICIAN ONLY menu (or 
any menu).  

For maximum accuracy, allow the sensor to warm 
up for 24 hours BEFORE setting the balance and 
preamp gain.  

 

SET BALANCE 

Attach a calibration cup to the sensor head and apply 
“zero gas” to the sensor and wait one minute for the 
reading to stabilize. Enter the TECHNICIAN ONLY 
menu and select the SET BALANCE option. Adjust 
the sensor output for a coarse zero. If the sensor is 
working properly, the reading prior to setting zero 
should be +/- 3% LEL.  

PREAMP GAIN SET

 

Apply a specified “span gas”  to the sensor and allow 
the GASMAX reading to stabilize (~ 60 seconds). 
Using the PreAmp Gain Set option, adjust the reading 
to a value within 5% of the span gas value (coarse 
adjustment only). Press NEXT to exit the menu.  

VII. CALIBRATION 