DU-AG-0018-25 Rev J 4/19/17 GCX Corp Page 2 of 8  

Height Locking Lever 

M6 X 12mm FHMS (3) 

Mounting Plate 

Table Mount 

*Supplied by Philips 

 

 

Warning!  

Do not stack the Table Mount and Spacer when 
installing the Table Mount. 

3.

  The combination of MP60/70 and an 8-slot or 4-slot FMS exceeds the acceptable load limit for the VHM Arm.  

MP60/70 can only be used without a FMS. 

 

4.

  If assistance is needed regarding an application, please contact a GCX product specialist at (800) 228-2555 or +1 707-

773-1100. 

 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
Attaching the Philips-supplied Table Mount to the Mounting Adapter

 

 

1.

  Fasten Philips-supplied Table Mount to Mounting Adapter with three (3) M6 x 12mm FHMS (see below). 

 
 
 
 
 
 

 

 
 
 
 
 
 

 

 

 

 

Mounting VHM Arm in Channel 

1.

  Thread the Height Locking Lever into the mounting hole in the right side 

of the Arm.  The Lever operates by turning clockwise to tighten or 
counterclockwise to loosen.  Ensure that the Locking Lever is in the 
locked position before mounting the Arm in the channel. 

 
 
 
 
 
 
 
 
 
 
 
 

Do not attach Table Mount if monitor 

has this mounting interface. 

Attaching the Philips-supplied Table Mount to the Mounting Plate 
 
Installation Note:
  Do not attach the Table Mount to the Arm/Mounting Adapter if 
you will be mounting a monitor with the mounting interface shown in photo (right). 
See "Mounting the Monitor" (page 3) for instructions specific to this type of 
mounting interface. 

 

Содержание VHM

Страница 1: ...ht of the mounted device The Arm housing should never be disassembled by non GCX personnel Additionally the mounted device should only be removed with the Arm in its highest vertical position and with...

Страница 2: ...Philips supplied Table Mount to Mounting Adapter with three 3 M6 x 12mm FHMS see below Mounting VHM Arm in Channel 1 Thread the Height Locking Lever into the mounting hole in the right side of the Ar...

Страница 3: ...itor in accordance with the Philips Table Mount instruction sheet CAUTION Ensure the device mounting locking mechanism is properly engaged at all times Fixed Mount 1 Fasten Spacer to Mounting Adapter...

Страница 4: ...or FMS 1 Align three 3 mounting holes in Lower Post Disc with three 3 threaded holes in Swivel Cup below left Using the 5 32 hex wrench provided fasten Post to Swivel Cup with three 3 10 32 x 3 8 SHCS...

Страница 5: ...ough the Adjuster Cover below center Lock Arm when Counterbalance Adjuster Bolt is accessible 3 With the 1 2 13mm socket wrench on the Adjuster Bolt carefully unlock and support the arm Turn the Adjus...

Страница 6: ...cup to gain access to the Swivel Tension Nut Tilting the Mounted Instrument and Adjusting Tilt Tension Installation Note Shown without mounted device for detail purposes 1 To tilt the instrument gras...

Страница 7: ...e necessary during vertical adjustment of mounted instrument see WARNING above Engagement of the Height Locking Lever is not required during normal operation of the counterbalanced instrument Position...

Страница 8: ...sive material to clean the mounting assembly 4 Damage caused by the use of unapproved substances or processes will not be covered by warranty We recommend testing of any cleaning solution on a small a...

Отзывы: