DU-AG-0021-100 Rev A 4/15/08 GCX Corp. Page 2 of 6
Attaching Top Plate/Channel to Top of Fabius GS
The following parts and hardware are included in this installation kit (hardware not shown):
Tools Required
• Drill (not provided)
• 1/4'' drill bit (not provided)
• Phillips screwdriver
1. Locate four (4) dimples in top of anesthesia machine. Dimples indicate locations of threaded inserts beneath top
surface. If dimples are difficult to locate, use Top Plate/Channel assembly as a template and mark top of machine
through four (4) mounting holes.
2. Using a 1/4'' drill bit, drill in locations marked in step 1 (drill only where dimples are present).
CAUTION:
Do not allow
drill bit to penetrate more than 1/4'' into top of machine. Drilling too deeply may damage threaded inserts and internal
components of the machine. Remove all drilling chips.
3. Align four (4) mounting holes in Top Plate with corresponding holes in top of machine (drilled in step 2). Fasten Top
Plate to machine with four (4) M5 x 30mm Pan Head Machine Screws (PHMS)
Item #
Description
Qty
1
Top Plate with 21'' Channel
1
2
M5 x 30 mm Pan Head Machine Screw (PHMS)
4
3
Label, Tip Hazard
1
Top Plate/Channel
M5 x 30mm PHMS (4)
Locations of Threaded Inserts (4)
in top of Fabius GS
1