To apply the cleaning solution do not spray it as the spray may enter into 

the inner parts of product and cause contamination or damage.
DO NOT use pressure wash as it could damage or contaminate the 

product.
If the inner parts of the product have been contaminated DO NOT 

continue to use the product under any circumstances. It must be 

withdrawn from service.

9. SERVICE, PRODUCT LIFE TIME AND        

  

MAINTENANCE

9.1. PRODUCT LIFE TIME

9.1.1. SERIAL NUMBER AND DATE OF PRODUCTION

An eight-figure serial number is stamped on valve body consists of the 

data as follows:

M RR XXXXX

M: month of manufacture

RR: year of manufacture

XXXXX: sequence number of product

Month:

A January

B February

C March

D April

E  May

F  June

G July

H August

I  September

J  October

K November

L  December

For example: Serial number A1000521 indicates the valve was manufac-

tured in 2010, In January, with sequence number 521.

9.1.2. PRODUCT LIFE TIME AND WASTE MANAGEMENT

Maximum life time of the product is 10 years from the manufacturing date. 

At the end of the  product’s life time (10 years maximum), the product must 

be withdrawn from service. The owner of the device shall prevent the reuse 

of the product and handle the product in compliance with “Directive of 

European Parliament and Council 2006/12/ES of 5th April 2006 on waste“.

Содержание ELITE ELH BS PROBE

Страница 1: ...GCE HEALTHCARE PICTU RE GCE HEALTHCARE ELITE OXYGEN CONSERVER MODEL ELH BS PROBE EN INSTRUCTION FOR USE...

Страница 2: ...in materials which do not normally burn in air DO NOT use oil wax polish or grease on any part of Sabre Elite as they may cause fire or explosion DO NOT use the unit if the next service due date on t...

Страница 3: ...STORAGE AND TRANSPORT CONDITIONS 20 60 C 30 60 C 10 95 20 70 600 1200 mbar 600 1200 mbar In case of storage at a temperature below 20 C or above 60 C do not operate the medical device until it has bee...

Страница 4: ...unit so the reservoir fills up to a level determined by the flow setting As an inhalation begins the small suction at the nose makes a valve open A timed pulse of oxygen is delivered from the reservo...

Страница 5: ...a danger of fire or explosion arising from oxygen pressure shocks Insufficient opening of the shut off valve could reduce actual flow delivered ATTACHING TO A BS 5682 SOCKET Insert the end of the BS...

Страница 6: ...patient Turn off at the cylinder shut off valve by turning the hand wheel in a clockwise direction Disconnect the tube from the flow outlet Refit pressure outlet and flow outlet protection cap Before...

Страница 7: ...data as follows M RR XXXXX M month of manufacture RR year of manufacture XXXXX sequence number of product Month A January B February C March D April E May F June G July H August I September J October...

Страница 8: ...son who also holds all necessary certificates in compliance with national standards for mounting and repair of dedicated gas devices For information about a service in your area please contact GCE or...

Страница 9: ...ose into unsorted municipal waste P4 Max outlet pressure closing pressure Q Outlet flow 11 WARRANTY The Standard Warranty period is two years from date of receipt by the GCE Customer or if this is not...

Страница 10: ...MANUFACTURER GCE s r o Tel 420 569 661 111 Zizkova 381 Fax 420 569 661 602 583 01 Chotebor http www gcegroup com Czech Republic GCE s r o...

Страница 11: ......

Страница 12: ...esk republika http www gcegroup com Doc Nr IFU0148 DOI 2020 08 26 Rev 00 TI A6 CB V1 UNITED STATES OF AMERICA CHINA INDIA RUSSIA MEXICO SWEDEN POLAND GERMANY FRANCE SPAIN ITALY IRELAND UNITED KINGDOM...

Отзывы: