20

 

NEOSERV 114

 

 

 gbo Medizintechnik AG 2004

 

Version 1.5

 

Notes in accordance with EC directive and Medical 
Devices Directive (MDD) 

 

The 

NEOSERV 114/114-H 

is a line-powered suction wave massage unit of protection class 

I

. Normally it 

serves as an suction application aid for current stimulation therapy. 
 
The device is in accordance with the EC directive for medical devices (93/42/EWG) and therefore carries 
the CE sign with the registration number of the notified body for medical devices. The according graphical 
symbol is placed on the type plate. 
 
According to the MDD, 

NEOSERV 114/114-H 

is a class 

IIa

 device. 

 
The manufacturer is only responsible for the safety, operational reliability and functionality of the device if: 

• 

the device is used in accordance with the instructions for use; 

• 

the electrical installation of the location where the device will be used meets the respective current 
requirements of electrical safety; 

• 

the device is not used in hazardous environments and humid locations; 

• 

mountings, amplifications, re-adjustments, modifications or repair works are carried out only by 
personnel authorized by the manufacturer; 

• 

the operator regulation of this EC directive is observed within the scope of MDD. 

 

Technical support may be obtained by the manufacturer, dealers or service authorized by the manufacturer. 
The product’s duration of life as scheduled by the manufacturer is 10 years. 
 

NEOSERV 114/114-H 

is an electronic device. For its disposal the according regulations for electronic 

devices have to be observed. Incidentals have to be disposed with residual waste. 
 
On request, the manufacturer will provide you with further technical descriptions for all repairable parts of 
the device, such as circuit diagrams, spare parts lists, and adjustment instructions as far as these are 
necessary for the qualified technical staff of the operator. 
 
 

Comments on electromagnetic compatibility (EMC) 

Medical, electrical devices are subject to special precautions concerning the EMC. They must be installed 
and operated according to the EMC-advice given in the accompanying documents. In particular medical, 
electrical devices may be influenced by portable and mobile RF-communication devices. 
 
The manufacturer guarantees the conformity of the unit with the EMC-requirements only when using 
accessories which are listed in the EC declaration of conformity. The usage of other accessories my cause an 
increased emission of electromagnetic disturbances or may lead to a reduced electromagnetic immunity. 
 
The unit must not be arranged physically close to other devices or stacked with other devices than 

DUODYNATOR 119

 or 

HiToP

®

. If such an order is necessary nevertheless, the unit must be observed in 

order to check it for the intentional operation. 
 
You find more EMC-comments in the chapter “Warnings and Safety Precautions” of this manual as well as 
in the Technical Information on the next two pages. 

Содержание NEOSERV 114

Страница 1: ...gbo Medizintechnik AG 2004 Version 1 5 Suction Wave Massage Unit NEOSERV 114 NEOSERV 114 H User Manual...

Страница 2: ...omissions All rights reserved No part of this manual may be reproduced in any form or by any means electronic mechanical or otherwise without the prior written permission of the gbo Medizintechnik AG...

Страница 3: ...Medizintechnik AG 2004 Version 1 5 3 Warnings and safety precautions Warning Warnings which have to be observed by all means Caution Observe the instructions for use Note Information that will facili...

Страница 4: ...nnector Unit 11 4 OPERATING THE UNIT 12 4 1 Switching on the Unit 12 4 2 Select Channels 12 4 3 Vacuum Setting 12 4 4 Pulses Setting 12 4 5 Applying the Suction Electrodes 13 4 6 Terminating the Treat...

Страница 5: ...nt selected is applied to the patient by means of the suction electrodes The NEOSERV 114 works as suction wave massage unit Liquids dirt or bacteria which are sucked up by the suction electrodes are c...

Страница 6: ...6 NEOSERV 114 gbo Medizintechnik AG 2004 Version 1 5 1 2 Short Instructions Figure 1 NEOSERV 114 Figure 2 NEOSERV 114 in combination with DUODYNATOR 119...

Страница 7: ...LED of the Standby key will be flashing 3 Connect all needed electrodes to the appropriate connectors by means of the electrode hoses red channel I green channel II Note If you use only 2 suction elec...

Страница 8: ...t session remove the suction electrodes from the skin and switch off all units or put them into standby mode 10 Clean the sponges on a daily basis 11 Observe the Operating Instructions relating to any...

Страница 9: ...ovided to the unit s connector socket power supply unit One end of each electrode hose is connected to the suction connectors obey colors Then connect the suction electrode to the other end of the fle...

Страница 10: ...rheating the pump will be automatically switched off and the LED will be flashing 3 4 Water Level If the water level within the water separator is too high the red water level indicator LED will be fl...

Страница 11: ...irtight 3 6 Vacuum and Pulse Regulator By means of the Vacuum regulator a continuously variable setting of the vacuum will be achieved The Pulse regulator is for setting the number of suction pulses p...

Страница 12: ...pressure points otherwise the suction connection is not airtight 4 3 Vacuum Setting The Vacuum regulator is used for selecting the desired vacuum pressure This vacuum pressure level can be set anywher...

Страница 13: ...in 5 Apply the electrodes to the skin and hold them in place until the sucking action sets in Note Each suction electrode can easily be taken off by raising the electrode e g for changing their the po...

Страница 14: ...ent sucking action Foreign particles have accumulated inside the suction electrode s Suction electrodes draw in secondary air Leakage in tubing electrode lines Controls front panel in lowest position...

Страница 15: ...user 6 1 Safety Controls Irrespective of the legal rules it is recommended to have the device checked by the manufacturer or by a service agency authorized by him at 12 months intervals The check shal...

Страница 16: ...ponges may impair the sponges conductivity To improve restore conductivity we recommend using a saline solution Remove the metal part simply by drawing off from the electrode Remove the rubber cover f...

Страница 17: ...hone or place them in a distance of 3 m from the device Cardiac pacemakers can extremely be disturbed In these cases the therapy should be only carried out under continuous pulse and ECG control If th...

Страница 18: ...nformity sign Caution Observe the instructions for use Application part ungrounded protection degree Type BF This product complies with WEEE Directive 2002 96 EG waste electrical and electronic equipm...

Страница 19: ...Relative humidity 30 75 Transport and storage Temperature range 5 C 50 C Relative humidity 90 none condensing gbo Medizintechnik AG reserves the right to modify the design and specification without p...

Страница 20: ...as scheduled by the manufacturer is 10 years NEOSERV 114 114 H is an electronic device For its disposal the according regulations for electronic devices have to be observed Incidentals have to be disp...

Страница 21: ...nity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete...

Страница 22: ...80 MHz to 800 MHz d 2 3 P for 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts according to the transmitter manufacturer and d is the recommended separation di...

Страница 23: ...NEOSERV 114 gbo Medizintechnik AG 2004 Version 1 5 23...

Страница 24: ...24 NEOSERV 114 gbo Medizintechnik AG 2004 Version 1 5 Art No 014 7 0054...

Отзывы: