Error messages
Error 1: current limit:
This error message appears, when a short circuit of the electrodes causes a
current which exceeds the maximum valid value.
Error 2: patient plug removed:
This error message appears, when the connector is unplugged during therapy
or the connector has a loose contact.
Error 3: current monitoring:
This error message appears, when contact problems in the cable system
causes a strong fluctuation of the current.
Error 4: open circuit:
This error message appears, when there is no current flow. This can be
caused by a cable defect or an error in the electrode placement.