| Using the Omni™ Blood Pressure Controller | User Guide |
13
PULSE MENU
Use this menu to set the heart rate. Adjust the
beats per minute using the
+
and
-
buttons. Press
OK
to save the changes and return to previous
screen.
AU. GAP MENU
Enable and set an interval of pressure where
Korotkoff sounds dissipate. The auscultatory gap
is disabled by default.
Pressing the ENABLE button activates the
UPPER and LOWER menus. Use these menus
to set the limits of the auscultatory gap.
Adjust the limits of the auscultatory gap using the
+
and
-
buttons. Press
OK
to save the changes
and return to previous screen. The default upper
and lower limits for the auscultatory gap are 100
mmHg and 90 mmHg, respectively.
VOLUME MENU
Increase or lower the volume of Korotkoff sounds
using the
VOLUME
menu. Press
OK
to save
the changes and return to previous screen. The
default value for the Korotkoff sounds volume is
5.
CALIB MENU
Calibrate the sensors inside the simulator
using this option. Set the standard against
which auscultation will be evaluated during the
exercise. Blood pressure cuff calibration should
be performed only when the Korotkoff sounds do
not match the systolic and diastolic values set
from Omni.
It is not necessary to re-calibrate Omni
for each training session. Omni will
remember previous calibration settings.
CALIBRATION INSTRUCTIONS
1. Assemble the system and turn Omni on.
2. Place the blood pressure cuff on the arm.
3. Connect the cuff’s extension to connector on
the control panel.
Содержание S415.100
Страница 2: ...2 User Guide Blood Pressure Training System with Speakers...
Страница 4: ...4 User Guide Care and Cautions Care and Cautions...
Страница 6: ...6 User Guide BP Arm Set Up BP Arm Set Up...
Страница 9: ...BP Arm Features User Guide 9 BP Arm Features...
Страница 11: ...Using the Omni Blood Pressure Controller User Guide 11 Using the Omni Blood Pressure Controller...
Страница 15: ...Appendix User Guide 15 Appendix...