background image

 

                                                    G-Finder Multi (O

2

 / CO / H

2

S / CH

4

) Portable Multi Gas Detector

 

 

 

GT-07919072501 

PAGE  12  of  14 

Rev 1.0(2020.05.18) 

 

 

 

Battery life decreases faster when the LCD 
back-light is turned on. If you do not need 
the LCD back-light, it is recommended to 
save battery by turning the LIGHT mode 
off. 

 

1)

 

Press the button twice to get into the menu. 
Then, "BUMP TEST" will be displayed on the 
screen. "LIGHT MODE" will be displayed on 
the screen by pressing the button six times. 
(or five times in case of G-Finder Multi 
without Bluetooth option) 

 

 

2)

 

Press and hold the button until 3 seconds 
countdown is displayed, then keep holding 
until the countdown is completed in order to 
perform the Light mode. 

 

 

3)

 

When the countdown is completed, “LIGHT-
ON” is displayed on the screen for a while then 
return to the Light mode menu. 

 

 

4)

 

In order to turn off the Light mode, proceed 
the above procedure ‘1)’. 

 

5)

 

Press and hold the button until 3 seconds 
countdown is displayed, then keep holding 
until the countdown is completed in order to 
turn off the Light mode. 

 

6)

 

When the countdown is completed, “LIGHT -
OFF” is displayed on the screen for a while 
then return to the Light mode menu. 

 

 

14.

 

Detection Range 

 

Please refer to the below table for the Detection 
ranges. 

Gas Type 

Detection Range 

O

2

 

0 to 25.0% Vol 

with 0.1 increments 

CO 

0 to 300 ppm 

with 1 increment 

H

2

0 to 100 ppm 

with 0.1 increments 

CH

4

 

0 to 100% LEL 

with 1 increment 

 

15.

 

Detectable Combustible Gases 

 

G-Finder Multi combustible gas sensor is 
optimized for methane detection. The sensor is 
able to detect and respond to the other 
combustible gases shown in the table below, but 
the accuracy of the measured gas concentration 
values may not be consistent. 

 

Combustible 

Gas 

Expected response 

at 20% LEL target gas 

Methane(CH

4

20 %LEL 

Propane(C

3

H

8

15 ~ 45 %LEL 

Butane(C

4

H

10

15 ~ 35 %LEL 

Pentane(C

5

H

12

13 ~ 45 %LEL 

Hexane(C

6

H

14

7 ~ 28 %LEL 

Ethane(C

2

H

6

42 ~ 50 %LEL 

Ethylene(C

2

H

4

4 ~ 5 %LEL 

 
 

 

Содержание G-Finder Multi

Страница 1: ... consequences for the environment 4 It is recommended performing a bump test prior to use G Finder Multi every day to confirm sensor response and alarm activation by exposing the detector to a concentration of target gas that exceeds the low alarm set point 5 For optimal performance please proceed zero calibration periodically 6 G Finder Multi is provided with anti static coating over the LCD wind...

Страница 2: ...5 RH IP IP 68 Sensor type O2 electrochemical cell CO electrochemical cell H2S electrochemical cell CH4 or C3H8 Non Dispersive Infrared Sensor C3H8 is an option in the future Alarms Visual vibrating audible min 95dB Display Liquid Crystal Display LCD Operating frequency 2402 MHz 2480 MHz RF output power 11 08 dBm Max RF Distance 7m at the non interference environment Battery Primary lithium thionyl...

Страница 3: ... blinks when the detector works normally without any functional errors Displays for remaining product life time Oxygen concentration is measured as percent by volume Combustible gas concentration is measured as lower explosive limit Toxic gas concentration is measured as parts per million Alarm event indicator This icon is displayed when an alarm event has occurred within the past 8 hours End of o...

Страница 4: ...will display current firmware version of the detector The number before the underscore _ indicates major number of the firmware version and the number after the underscore indicates minor number In other word 1_00 indicates the firmware number 1 00 Firmware Version 3 By pressing the button once from the above phase it will display the remaining product life The remaining life will be counted autom...

Страница 5: ...s rapidly when the detector is in alarm condition For the gas type O2 a low alarm occurs when the measured concentration value is lower than the low alarm setting value while a high alarm occurs when the measured concentration value is higher than the high alarm setting value The following alarm settings are default for each gas type Gas Type Low Alarm High Alarm O2 19 5 v v O2 23 5 v v O2 CO 30 p...

Страница 6: ...n of all 4 types for the bump test in 60 seconds the bump test process is performed automatically The result will be displayed on the screen at the end of the test If an alarm occurs during the bump test the test is succeeded otherwise the test is failed If the test is failed icons are appeared additionally with the bump test mode display on the screen 8 Zeroing the Sensor In case of O2 sensor ove...

Страница 7: ... keep holding until the countdown is completed in order to perform the selected zero calibration 7 Then the zeroing process will be performed automatically Wait until the zeroing process is completed 8 When the zeroing process is completed PASS or FAIL appears 9 After displaying the result ZERO TEST is displayed on the screen again automatically 10 If FAIL appears repeat the zeroing process accord...

Страница 8: ... is displayed on the screen If the button is pressed only once it will skip to the next menu 5 If you enter selecting menu for the gas calibration by pressing the button twice when the SPAN TEST is displayed on the screen ALL SPAN is displayed first and the display changes to O2 SPAN COMB SPAN H2S SPAN and CO SPAN sequentially by pressing the button 6 When the corresponding gas calibration menu di...

Страница 9: ...ODE will be displayed on the screen by pressing the button three times 2 Press and hold the button until 3 seconds countdown is displayed then keep holding until the countdown is completed in order to perform the IRDA communication mode 3 Then the display on the screen is changed to I ON which indicates that the IRDA communication mode is ready for normal operation While I ON is displayed on the s...

Страница 10: ...n 2 Press and hold the button until 3 seconds countdown is displayed then keep holding until the countdown is completed in order to enable the connection of Bluetooth communication mode 3 Then ON XX XX XX XX is displayed on the screen and the icon blinks to wait for a connection from the Bluetooth device 4 Run the application on the Bluetooth device search and select GFM 400 which has the same MAC...

Страница 11: ...tdown is displayed then keep holding until the countdown is completed in order to perform the Silent mode 3 When countdown is completed SILENT ON is displayed on the screen for a while then return to the Silent mode menu with icon After this moment this Silent mode is turned on with icon lights up continuously 4 In order to turn off the Silent mode proceed the above procedure 1 until SILENT MODE d...

Страница 12: ...he above procedure 1 5 Press and hold the button until 3 seconds countdown is displayed then keep holding until the countdown is completed in order to turn off the Light mode 6 When the countdown is completed LIGHT OFF is displayed on the screen for a while then return to the Light mode menu 14 Detection Range Please refer to the below table for the Detection ranges Gas Type Detection Range O2 0 t...

Страница 13: ... NEPSI GYJ19 XXXX KC R C G99 GFM 400 2 The product is in conformity with the following standards IECEx IEC 60079 0 2017 IEC 60079 11 2011 KCs Announcement No 2019 15 of Ministry of Employment and Labor ATEX EN IEC 60079 0 2018 EN 60079 11 2012 NEPSI GB 3836 1 2010 GB 3836 4 2010 GB 3836 20 2010 3 The product may be used in zones 0 1 2 with flammable gases and vapors with apparatus groups IIC and w...

Страница 14: ...e Target Gas Calibration Gas Measurement range vol Default blank CH4 CH4 0 5 MM2 5 CH4 CH4 0 2 5 MM100 CH4 CH4 0 100 MP1 5 CH4 C3H8 0 1 5 MP2 5 CH4 C3H8 0 2 5 PP1 5 C3H8 C3H8 0 1 5 PP2 5 C3H8 C3H8 0 2 5 Y Housing body color Blank Orange default YE yellow GN green VT violet BK black etc Y Option which does not affect intrinsic safety This option code is needed at the order phrase for different hous...

Отзывы: