Carefully place the metallic cover back over the seated Containment Window – ensure that
the cover is fully seated (note that the Containment Window metallic cover can only be
assembled in one orientation).
•
Lastly, return to the software interface and press the ‘Maintenance Complete’ button.
•
The full Sample Receptacle will automatically re-insert itself back into the ejection position.
•
Replace the Sample Tray, and return to the ‘Welcome’ screen for normal operation.
Figure 6-2: Maintenance – Holder Receptacle Position (Left) Containment Window Assembly (Right)
6.1.2 Cleaning the Containment Window
If the Containment Window is to be cleaned (not replaced), then the following procedure must be
followed. Note that gloves are required to complete this procedure and the User must refer to
Figures 6-1, 6-2 and 6-3 to supplement the procedure below.
•
In the software interface, navigate to the ‘Welcome’ screen and enter the assigned
‘Maintainer’ pass code (refer to Section 5.1, the Maintainer pass code is provided during
training).
•
Once in ‘Maintenance’ mode, press the ‘Conduct Maintenance’ button near the middle of the
screen.
•
The Sample Receptacle will automatically displace further outside of the unit providing the
required access to the Containment Window Assembly.
•
Remove the Sample Tray, and set it aside.