background image

Gastops Ltd. 

 

 

C009423, Rev. 4 

 

page 44 

All use of this publication and ChipCHECK by user shall be subject to the restrictions and conditions as set out in this publication. 

 

Carefully place the metallic cover back over the seated Containment Window – ensure that 
the cover is fully seated (note that the Containment Window metallic cover can only be 
assembled in one orientation). 

 

Lastly, return to the software interface and press the ‘Maintenance Complete’ button. 

 

The full Sample Receptacle will automatically re-insert itself back into the ejection position. 

 

Replace the Sample Tray, and return to the ‘Welcome’ screen for normal operation. 

 

Figure 6-2:  Maintenance – Holder Receptacle Position (Left) Containment Window Assembly (Right) 

6.1.2 Cleaning the Containment Window  

If the Containment Window is to be cleaned (not replaced), then the following procedure must be 
followed.  Note that gloves are required to complete this procedure and the User  must refer to  
Figures 6-1, 6-2 and 6-3 to supplement the procedure below. 

 

In the software interface, navigate to the ‘Welcome’ screen and enter the assigned 
‘Maintainer’ pass code (refer to Section 5.1, the Maintainer pass code is provided during 
training). 

 

Once in ‘Maintenance’ mode, press the ‘Conduct Maintenance’ button near the middle of the 
screen. 

 

The Sample Receptacle will automatically displace further outside of the unit providing the 
required access to the Containment Window Assembly. 

 

Remove the Sample Tray, and set it aside. 

Содержание B166500

Страница 1: ...ESIGNS DISCLOSED HEREIN ARE THE EXCLUSIVE PROPERTY OF GASTOPS LTD AND ARE NOT TO BE USED OR DISCLOSED TO ANY OTHERS WITHOUT THE WRITTEN CONSENT OF Gastops LTD ANY RECIPIENT BY ITS RETENTION AND USE AGREES TO HOLD IN CONFIDENCE ANY INFORMATION TECHNICAL DATA AND DESIGNS DISCLOSED HEREIN THE FOREGOING SHALL NOT APPLY TO PERSONS HAVING PROPRIETARY RIGHTS TO SUCH INFORMATION TECHNICAL DATA OR DESIGNS ...

Страница 2: ... unit requires service then it shall only be serviced or repaired by Gastops authorized technicians For the maintenance procedures discussed herein there is no need to remove any covers or modify the unit The User is strictly forbidden from modifying the unit removing any covers or to alter interfere modify remove circumvent defeat or de active any safety system or interlocks refer to Section 2 2 ...

Страница 3: ...ing Once in Maintenance mode press the Conduct Maintenance button near the middle of the screen The Sample Receptacle will automatically displace further outside of the unit providing the required access to the Containment Window Assembly Remove the Sample Tray and set it aside Remove the Containment Window from its housing by carefully lifting the metallic cover located on top of the Sample Recep...

Страница 4: ...on Figure 6 2 Maintenance Holder Receptacle Position Left Containment Window Assembly Right 6 1 2 Cleaning the Containment Window If the Containment Window is to be cleaned not replaced then the following procedure must be followed Note that gloves are required to complete this procedure and the User must refer to Figures 6 1 6 2 and 6 3 to supplement the procedure below In the software interface ...

Страница 5: ...owly drag the Isopropyl soaked cloth over the surface of the Containment Window pulling parallel to the exposed surface of the Containment Window ensuring no pressure is applied to the window itself Repeat these steps as required Once the Containment Window is clean place it back onto the Sample Receptacle ensuring that the Containment Window is properly seated on the centering features of the Rec...

Страница 6: ...ry Module Follow the set up instructions as described in Section 3 3 Figure 6 3 Power Entry Module and Main Power Switch 6 2 Unit Service No Allowable Service In the event of any equipment failure or component replacement Users are strictly forbidden to perform any kind of service There is absolutely NO SERVICE ALLOWED by the User Service activities are restricted to only Gastops authorized techni...

Страница 7: ...r designated customer support representative Refer to Section 8 0 Table 6 1 Basic Trouble Shooting Procedures Symptom Possible Remedy Software interface does not respond to user input Conduct a complete unit shutdown and power down Then power up and restart the unit as per Section 3 3 If this does not work then contact Gastops Customer Support See Section 8 0 Error occurs in software that ceases u...

Страница 8: ...ion 5 4 The unit fails both Standardization and Calibration Power cycle the unit Perform the maintenance sequences defined in Section 6 0 Repeat Calibration and the complete Standardization If the unit still fails Calibration then contact Gastops Customer Support See Section 8 0 Chip Analysis No Particles Found result after analysis Verify that the correct Sample Patch was inserted into the unit f...

Страница 9: ... required Please call customer support Conduct a complete unit shutdown and power down Then power up and restart the unit as per Section 3 3 If this does not work then contact Gastops Customer Support See Section 8 0 Sequence contains no matching element Object reference not set to instance of an object Out of memory A generic error occurred in GDI Conduct a complete unit shutdown and power down T...

Страница 10: ...removable foam insert protects the front face of the ChipCHECK unit When unboxing the unit this insert must be removed before the ChipCHECK can be removed from the transport case To remove the unit from its transport case simply open the case lid lift out the foam insert and then lift the unit out of the box see Figure 7 2 NOTE Slots for storing accessories can be found in the foam insert and in t...

Страница 11: ...ed custom foam cut out Note that there is only one proper orientation for the unit to be placed otherwise the Transport Case will not close properly 6 Replace the instrument s accessory foam insert and place all of the instrument s accessories and power cable into their respective custom foam cut outs 7 Replace the Transport Case cover and re fasten all latches 8 Lock or fasten each latch where po...

Страница 12: ...tions Foam insert removed Figure 7 4 Foam Insert Storage Locations Space for Operator s Manual located beneath ChipCHECK unit Storage slots beneath foam insert for Sample Cases Calibration Standardization Patches Sample Tray External Keyboard optional External USB Hub optional External Mouse optional Wash Bottle optional Sample Preparation Tools optional Sample Patches Ethernet Adapter Tweezers op...

Страница 13: ... 200x Sample Patches 1x Standardization Standard 1x Calibration Standard 20X Sample Patch Cases 1x USB Ethernet Adapter Additional optional accessories may include the following External keyboard External mouse External USB hub Sample preparation tools tweezers probe set wash bottle 7 2 Storage When storing the unit temporarily take note of the environmental limitations defined at the beginning of...

Страница 14: ...e restrictions and conditions as set out in this publication 8 0 Customer Support Gastops technical support representatives for ChipCHECK can be reached at Phone 1 613 744 3530 Fax 1 613 744 8846 Toll Free 1 800 363 8658 Email support gastops com Gastops Limited 1011 Polytek Street Ottawa ON K1J 9J3 Canada ...

Отзывы: