GC-IMS
– User Manual
115/214
7.5 Prepare the device for operation
Before using the device for the first time or after being disconnected from the
nitrogen source for some time it must be cleaned to ensure proper operating
conditions. In this case start the
cleaning mode
.
An appropriate cleaning period duration must be chosen depending on how long the
device was switched off and on the extent of contamination. It is recommended to
clean the device for at least
15 hours
before operating it for the first time. If the
Spectrum is not clean repeat the process. A
reference spectrum
for the evaluation
can be found in the supplied
Analytical Approval
.
INFORMATION!
After the cleaning process the device needs at least
2 hours
to cool
down the internal components to their required temperatures.
The duration of the cool down process is depending on the
temperature default settings.
1
Inspect spectrum for
contamination. Control
the baseline. Start
cleaning when
contaminated.
Contamination is
indicated by peaks or the
disappearing of the RIP.
Содержание GC-IMS
Страница 1: ...User Manual GC IMS G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 139: ...GC IMS User Manual 139 214 3 Wait until the exporting process is completed The measurement files has been downloaded...
Страница 150: ...GC IMS User Manual 150 214 4 Device without top cover...
Страница 157: ...GC IMS User Manual 157 214 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 164: ...GC IMS User Manual 164 214 17 Tighten the three screws of the frontplate...
Страница 188: ...GC IMS User Manual 188 214 9 4 Technical drawing Internal Gasflow...
Страница 212: ...GC IMS User Manual 212 214 Tutorials VOCal Manuals VOCal...