MGC Pump User’s Manual
v1.0
2
Portable gas detectors you can count on.
RI
Once the screen displays “APPLY GAS”, apply gas to the detector’s gas inlet. As the detector
detects gas, the sensor readings will be displayed and adjusted throughout the calibration.
Once calibration is complete
the detector will display the next calibration date before
returning to normal (alarming) operation. If a sensor fails to calibrate, the detector will
display an error for the failed sensor. Check your gas connections and concentration before
attempting a second calibration. If a sensor fails to calibrate after a second attempt, contact
GCT for either warranty replacement or replacement sensor(s).
Battery
The battery is displayed as both a percentage and a 3-bar battery icon. The percentage
calculation is approximate and can be used to provide a rough estimation of the time
remaining.
When the detector determines that only 20 minutes of run time remains, it will beep, flash
and display “LOW BAT” on the screen. This will repeat with 10 minutes remaining, and then
continuously for the last 5 minutes of run time. When the battery has expired, the detector
will give 5 long beeps and flashes, display “OFF” and then shut down.
To charge the detector, connect it to the provided charging adapter and plug it into an AC
outlet. The detector will cycle the battery icons in a charging fashion until the battery is fully
charged.
DO NOT charge the instrument in a combustible atmosphere.
DO NOT charge the instrument in temperatures above or below the specified range of 0°C to 40°C
DO NOT substitute any other battery type than specified and supplied by Gas Clip Technologies.