Multi Gas Clip User’s Manual
UM MGC V2.0
8
3 of 21
WARNING STATEMENTS/AVERTISSEEMENT
Do not substitute components as this may interfere with the intrinsic safety of the device .
Ne remplacez pas les composants car cela pourrait interférer avec la sécurité intrinsèque de l’appareil .
DO NOT substitute any other battery type than specified and supplied by Gas Clip Technologies .
Only use Gas Clip Technologies chargers and parts in the detector . Nonapproved parts will void the warranty and are
considered unsafe .
Before each use, check that all sensor and alarm ports are clear of any obstructions, i .e . debris or blockage .
The detector contains a lithium battery that must be disposed of by a qualified recycler . Check local regulations for
proper disposal .
DO NOT charge the instrument in a hazardous location .
DO NOT use IR communications when an explosive atmosphere may be present .
If you suspect any malfunction or have any technical problems, contact GCT at 1-877-525-0808 .
The battery may present a fire or chemical burn hazard if mistreated . Do not disassemble, heat above 100°C (212°F),
or incinerate . Contact Gas Clip Technologies for replacement instructions . Use of another battery may present a risk of
fire or explosion .
DO NOT charge the instrument in temperatures above or below the specified range of 0°C to 40°C .
Keep new and used batteries away from children .
DO NOT expose the detector to known combustible sensor poisons such as, but not limited to, alcohol and citrus
based cleaners, silicones, lead compounds (e .g . tetraethyl lead), sulfur compounds, phosphorus, halogenated
hydrocarbons and aerosols . Exposure to poisons may impair the accuracy and/or response time of the detector . This
applies to pellistor version only .
If suspected combustible sensor poisoning has occurred, recheck the instrument (both calibrate and bump test) .
The detector should be bump tested before use with a known concentration of gas to confirm its ability to respond to
gas . Calibrate the detector if the readings are not within the specified limits .
Le détecteur doit être testé avant utilisation bosse avec une concentration connue de gaz pour confirmer sa capacité
à répondre à gaz . Calibrer le détecteur si les relevés ne sont pas dans les limites spécifiées .
Any rapid up-scale reading, followed by a declining or erratic reading, may indicate a gas concentration beyond upper
scale limit which may be hazardous .
Toute rapide haut de gamme lecture suivie d’une lecture erratique baisse ou peut indiquer une concentration de gaz
supérieure à la limite supérieure de l’échelle qui peut être dangereux .
Strong Electromagnetic Interference (EMI) may cause incorrect operations .
Only the combustible gas detection portion of this instrument has been assessed for performance by CSA
International .
READ FIRST BEFORE OPERATION
Gas Clip Technologies (GCT) Multi Gas Clip (MGC) detectors are personal safety devices designed to detect the presence
of specific toxic gases: carbon monoxide (CO), hydrogen sulfide (H
2
S), oxygen (O
2
) and combustible gases/Lower
Explosive Limit (LEL) . Before operation, ensure that you have been properly trained on the use of the equipment and the
appropriate actions to take in the event of an alarm condition .