background image

Multi Gas Clip

Infrared

Multi Gas Clip 

Pellistor

G

A

S

C

L

I

P

T

E

C

H

.

C O

M

User’s Manual

Содержание Multi Gas Clip Infrared

Страница 1: ...Multi Gas Clip Infrared Multi Gas Clip Pellistor G A S C L I P T E C H C O M User s Manual...

Страница 2: ...play Details 6 Alarms 7 Default Alarms 7 Alarm Behavior 7 Detector Menus Options 8 Main Menu 8 Status Menu 8 Option Menu 9 Adjustable Options 9 Detector Maintenance 11 Bump Testing 11 Manual Bump Test...

Страница 3: ...cones lead compounds e g tetraethyl lead sulfur compounds phosphorus halogenated hydrocarbons and aerosols Exposure to poisons may impair the accuracy and or response time of the detector This applies...

Страница 4: ...nance LEDs 3 Display 4 Audible Alarm Port 5 Power Menu Button 6 Dual Tox Sensor Port 7 LEL Sensor Port 8 O2 Sensor Port 9 Calibration Cap Ports 10 Infrared IR Communication Window 11 Alligator Clip wi...

Страница 5: ...LOW BAT and then turn off again Once activated the detector will begin to display gas readings immediately Each sensor will show a chasing 0 for the sensor reading while it is stabilizing and being s...

Страница 6: ...ime to apply gas The Battery Charge Level 3 is displayed in a 3 bar battery icon as well as a percentage The percentage calculation is approximate and can be used to provide a rough estimation of the...

Страница 7: ...on fast LED flash Low Battery 20 minutes remaining single beep flash battery icon on 10 minutes remaining single beep flash battery icon flashing 5 minutes remaining continuous beep flash every 5 seco...

Страница 8: ...lay in the following order 1 L CAL the last calibration date 2 CAL DUE or CAL IN ___ d denotes if a calibration is due or how many days until it is due 3 L BUMP the last bump test date 4 BUMP DUE or B...

Страница 9: ...the detector when to go into alarm Alarm limits may be disabled by setting them to zero Caution Confirm alarm levels with local laws regulations before operation SAFE Display SAFE will be displayed i...

Страница 10: ...LEL typically 2 5 in North America and 2 2 in Europe Language The detector will display all its text prompts in any of six languages English German French Spanish Italian or Portuguese Note OL Err and...

Страница 11: ...en show in the following order 1 Current Date Time DD MON TT TT below stand for Day Month Time 2 User Programmed Text Message 3 SHOW STATUS Prompt When the screen displays SHOW STATUS quickly press an...

Страница 12: ...eturn to normal operation after a short period of time The bump due date will be automatically reset based on the new bump date The bump test will be recorded in the Bump Log When a bump test interval...

Страница 13: ...signate which sensor did not pass The failed sensor will be disabled and will need to be replaced see Err in Failures FAQ s section During a manual bump test the audible visual vibrating prompting to...

Страница 14: ...d down the power menu button through the shutdown sequence see Turning Off the Detector section and the calibration countdown sequence that follows it Display will show During these two countdowns the...

Страница 15: ...MGC Vehicle Charger MGC Charging Station or MGC Multi Charger all authorized GCT Multi Gas Clip Optional Accessories The detector will cycle the battery icons in a charging fashion until the battery i...

Страница 16: ...ump tests are differentiated from normal events when the alarm condition occurs inside of an MGC Dock or MGC Wall Mount Dock or when the detector is manually bump tested according to the Bump Testing...

Страница 17: ...Air Stone Particulate Filter MGC Calibration Cap 10 ft Sampling Hose and Hand Aspirator Pump Assembly with Quick Connect hose 3 ft Calibration Test Hose with Quick Connect 0 5Lpm Regulator with Quick...

Страница 18: ...d availability FAILURES FAQ S Err If a sensor is displaying Err this sensor has failed and is therefore disabled Contact GCT for either warranty replacement or replacement sensor s BUMP DUE If the det...

Страница 19: ...n activation Sensors battery and circuitry tests run continuously Ingress Protection IP 67 Warranty Full 2 years Gas Range Resolution Accuracy T90 Gases H2S 0 100 ppm 0 1 2 ppm 30s CO 0 500 ppm 1 5 pp...

Страница 20: ...ent has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful int...

Страница 21: ...to repair of the product by any person other than the authorized dealer or installation of unapproved parts on the product The obligations set forth in this warranty are conditional on 1 Proper storag...