Declaration - Electromagnetic Emissions and Immunity
For equipment and systems that are not life-supporting and are specified for use only in a shielded location.
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result
in improper operation. If such use is necessary, this equipment and the other equipment should be observed
to verify that they are operating normally. Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer than 30 cm (12 in.) to any part of the unit,
including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment
could result.
The Index BPM device is intended for use in the electromagnetic environment specified in the following table.
The customer or the user of the Index BPM device should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment -
Guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
Contact: ±8 kV
Air: ±2 kV, ±4 kV,
±8 kV, ±15 kV
Contact: ±8 kV
Air: ±2 kV, ±4 kV, ±8 kV,
±15 kV
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
N/A
1
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
+ 0.5kV, +1kV line(s) to
line(s)
+0.5kV, +1kV,+ 2kV
line(s) to earth
N/A
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short inter
ruptions, and voltage
variations on power
supply input lines
IEC 61000-4-11
Voltage dips:
0 % UT; 0,5 cycle
0 % UT; 1 cycle
70 % UT; 25/30 cycles
Voltage interruptions:
0 % UT; 250/300 cycles
Voltage dips: N/A
Mains power quality should
be that of a typical commer
cial or hospital environment.
If the user of the Index BPM
requires continued operation
during power mains interrup
tions, it is recommended that the
Index BPM be powered from an
uninterruptible power supply or a
battery.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz and 60 Hz
The Index BPM power frequency
magnetic fields should be at
levels characteristic of a typical
location in a typical commercial
or hospital environment.
The Index BPM device is intended for use in the electromagnetic environment specified in the following table.
The customer or the user of the Index BPM device should assure that it is used in such an environment.
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
1
Not Applicable. The Index BPM is battery-powered, not AC powered.