Operation
Operation manual VCpro 600
45
6.5
Solvents table
OPERATION MODE
►
The selection of the parameter pressure via the solvent table is only
possibly in the operation modes Manual and Program.
1. On the navigation bar, select the operation mode by encoder.
2. Select the parameter Set Pressure at the current operating mode.
3. Press the
PMIN
button on the button bar. The table for selection of the pressure parame-
ters appears:
Name
CAS No.
Vapor pressure p [mbar] at 40 °C
1,1,2,2,-TeCA
630-20-6
35
1,2,-dichlorethane
107-06-2
210
1,2,-dichlorethylene (c)
156-59-2
479
1,2,-dichlorethylene (t)
156-60-5
751
2-methyl-2-propanol
75-65-0
130
3-methyl-1-butanol
123-51-3
14
Acetone
67-64-1
556
Benzene
71-43-2
236
Chlorobenzene
108-90-7
36
Chloroform
67-66-3
474
Cyclohexane
110-82-7
235
Diisopropyl ether
108-20-3
375
Dioxan
123-91-1
107
DMF
68-12-2
11
Ethanol
64-17-5
175
Ethyl acetate
141-78-6
240
Ethyl methyl ketone
78-93-3
243
Heptane
142-82-5
120
Hexane
110-54-3
335
Isopropyl alcohol
67-63-0
137
Methanol
67-56-1
337
Methyl chloroform
71-55-6
300
n-butanol
71-36-3
25
n-pentanol
71-41-0
11
n-propyl alcohol
71-23-8
67
Pentachlorethane
76-01-7
13
p-Xylole
106-42-3
25
Tetrachlorethylene
127-18-4
53
Tetrachlormethane
56-23-5
271
Tetrahydrofuran
109-99-9
402
Toluene
108-88-3
77
Trichlorethylene
79-01-6
183
Water
7732-18-5
72