Technical Data
115051
19
4.3
Device data
4.3.1
Laboratory-Vacuum-Systems
Parameter
Unit
LVS
600 T
601 T
602 T
610 T
611 T
Pumping speed 50/60 Hz
DIN 28432
at speed of 1500 rpm
m
3
/ h
4.5 / 4.9
Ultimate pressure
at speed of 1500 rpm
mbar
2
Ultimate pressure
with gas ballast
at speed of 1500 rpm
9
Max. inlet pressure
bar
1
Max. outlet pressure
1
Intake and pressure ports
-
Hose nozzle DN 8
for hose inside diameter 8 mm
Ambient temperature
°C
+ 10 to + 40
Max. Operating gas
temperature
+
4
0
Bearing
-
maintenance-free
Reference surface
sound pressure level
DIN EN ISO 2151
dB (A)
44
Voltage / Frequency
(different data upon customer
request)
V, Hz
230, 50/60 (115, 50/60)
(generally with motor protection switch, switch and cable)
Motor power
W
370
Operating mode
-
S 1
Motor / Type of protection
DIN EN 60529
IP 54
Motor / Class of insulation
DIN EN 600034-1
F (160°C)
Type Examination
Certificate no.
-
WELCH_ATEX_03-01
Designation EX
II3G IIC T3 X (internal Atm. only)
Dimensions (W/D/H)
mm
360 / 310 /
395
360 / 310 / 445
Weight
kg
23.2
23.4
23.2
23.4
23.2
Order numbers for :
-
230 V
inclusive mains connection cable
IEC with plug CEE, UK
-
LVS …
-
LVS … with dial gauge digital
-
LVS … netvac 1
115051
-
-
115057
115057-10
11505
7
-02
115053
-
-
115054
-
115054-03
115055
-
-
115 V
inclusive mains connection cable
IEC with plug US
-
LVS
…
-
LVS … with dial gauge digital
-
LVS … netvac 1
115051-01
-
-
115057-01
115057-11
11505
7
-03
115053-01
-
-
115054-01
-
-
115055-01
-
-
100 V
inclusive mains connection cable
IEC with plug J
-
LVS …
-
LVS … with dial gauge analog
115051-05
-
115057-05
115057-12
115053-05
-
115054-05
-
115055-05
-