Elimination of operating faults
Operation Manual LVS 105 T
– 10 ef
+
53
Type of error
Cause
Measure
Au-
tho-
rized
It is not possible to
set the parameter
Firmware error
Upload the current firmware
1
Faulty electronics
Repair the device
3
Vacuum controller
cannot be set or
programmed
Faulty electron-
ics/display
Repair or replace
3
Incorrect language
System setting set
incorrectly
Set the language: first option under
system settings: see section
1
10
– Reset to fac-
tory settings
1
Access locked
ADMIN
or
USER
ADMIN or possibly
USER password is
set
Have unlocked by authorized person;
see sections
(1), 2
No access to vac-
uum controller via
PC
No connection to PC Check cable connection of the data
connection and check connection
status under
System setting
Net-
work
Status "connected"
1
IP addressing not
correct
Check the IP address; if necessary,
set DHCP to "dynamic", see section
7
– Network
1
Incorrect actual
value of pressure
Wrong calibration
Recalibrate the pressure sensor;
see section
– Control systems/calibration
1
Reset the device; see section
– Reset to fac-
tory settings
1
REPAIR
►
Repairs may only be performed by specialists or by the manufacturer;
see section 8 Maintenance.