SECTION 3: OPERATION

3.10  Starting Procedures

3.11

  Starting a Welch Aspiration/Filtration Vacuum System

 

Before attaching the pump to a system, familiarize yourself with the function and action of the 

 

Aspiration/Filtration Vacuum System that you have acquired. Review the power requirements as 

 

described in Section 2.60. Welch recommends running the pump for a few minutes to warm it up, before

 

use. The warm-up improves the pumps ability to handle humid air. Set the device on a level, sturdy 

 

surface so the controls can be easily reached and adjusted. Connect one end of the application tubing

 

to the lid vacuum port.

 

Note:  The in-line hydrophobic filter is designed to prevent foreign material from being 

 

 

drawn into the device. Pump performance may decrease if the hydrophobic filter gets wet or

 

 

build up of forgein materialsor Medfilter occurs. It may be necessary to periodically change 

 

 

the hydrophobic filter 

 

CAUTION

Do not move this device with a full collection jar.

 

1. 

Be sure that the power switch is in the OFF (down) position. Plug in the device and turn the 

 

 

switch to the ON (up) position.

 

2. 

Block the application tube leading from the collection jar and allow the gauge to reach stable 

 

 

vacuum level.

 

3. 

Set vacuum level by turning the regulator knob counter-clockwise to increase vacuum or 

 

 

clockwise to decrease vacuum..

 

Note:  This procedure should be performed each time you use this device.

3.12 

   Cleaning The Device

 

1. 

Make sure the ON/OFF switch is in the OFF (down) position.

 

2. 

Unplug the device from the wall outlet and allow suction level to drop.

 

3. 

Remove the application tube from the lid vacuum port. Tubing should be cleaned after every use. 

 

 

Run hot water through it. Soak in a solution of one part of vinegar to three parts hot water 

 

 

for 20 minutes. Rinse with hot tap water and air dry.

 

4. 

Remove the pump tubing from the top of the jar lid.

 

5. 

Remove the jar from the holder and place on a firm, flat surface. Release the tabs and lift the lid 

 

 

from the jar.

 

6. 

Wash the jar, gasket, lid and tube in a hot water/dishwashing detergeant solution and rinse with

 

 

hot tap water. Soak in a solution of one part of vinegar to three parts hot water for 20 minutes. 

 

 

Rinse with hot tap water and air dry.

 

7. 

Wipe the outside of the case with a clean, damp cloth after each use.

3.20  Leak Detection

3.21

  The importance of eliminating all leaks in a vacuum system is obvious. The pump must remove this 

 

added volume of leaked gas to maintain the desired vacuum. Leaks for these pumps can be located by 

 

slightly pressuring the system and painting the suspected area with a thick soap solution. Escaping air 

 

will produce soap bubbles.

8

Содержание WELCH 2511

Страница 1: ...76 8800 Fax 847 677 8606 Technical Support E Mail gdwelchvacuum gardnerdenver com Web Page www welchvacuum com Printed in USA Part No 642783 Approval Agency s Warning Not recommended for pumping acid...

Страница 2: ...e failure to observe the instruction could result in damage to the equipment associated equipment and process These units confirm to the SI International system of units of measurement The following s...

Страница 3: ...uum Connection 2 80 Vacuum Gauges 2 90 Traps Section 03 OPERATION 3 10 Starting Procedures 3 20 Leak Detection 3 30 Operating Pressure Range 3 40 Shutdown Procedures Section 04 MAINTENANCE 4 10 Mainte...

Страница 4: ...etermining maximum air pressure for that tool or attachment applies to model 2511 only Be sure to properly identify intake and discharge before using pump See Section 2 50 WARNING Do not fill the coll...

Страница 5: ...s involving a semi flexible surface such as a bench top they help to isolate noise and eliminate creeping All Aspiration Filtration Vacuum Systems should be mounted on a horizontal plane 2 40 Pump Loc...

Страница 6: ...rge pressure to be set between atmospheric and the maximum possible for the pump model Model 2511 is equipped with a pressure port Note Adjusting the vacuum regulators will affect the performance of t...

Страница 7: ...and off until the fault is corrected 2 63 Classification And Symbol Glossary 2 70 Vacuum Connection 2 71 For best results Welch recommends the length of the tubing between the pump and the chamber be...

Страница 8: ...t vacuum level by turning the regulator knob counter clockwise to increase vacuum or clockwise to decrease vacuum Note This procedure should be performed each time you use this device 3 12 Cleaning Th...

Страница 9: ...e The pump employs a non lube piston and cylinder No maintenance is necessary for the bearings All bearings are sealed and permanently lubricated Lubrication should not be attempted The units are buil...

Страница 10: ...See Section 2 80 5 20 PRESSURE PROBLEMS 5 21 Leakage and contamination are the general causes of problems associated with poor pressure To operate at maximum efficiency a system must be thoroughly cl...

Страница 11: ...2 Defective switch 2 Contact your service center or distributor See back page 3 Vacuum still exists 3 Disconnect tubing from the in system collection jar to release vacuum Reconnect tubing and turn o...

Страница 12: ...5 2515C 75 2515C 76 Free Air Displacement CFM L min 1 2 34 1 0 28 1 2 34 Maximum Vacuum 27 2 in Hg Ultimate Vacuum 70 Torr 93 mbar Dimensions L x W x H in cm 14 8 x 8 3 x 10 Inches 37 5 x 21 x 25 5 cm...

Страница 13: ...at have been used for any period however short will be repaired under this warranty rather than replaced The warranty is effective for one year from the date of original purchase when 1 The warranty c...

Страница 14: ...8 11 Model 2515B 75 2515C 75 2515C 76 COLLECTION JAR LONG TUBING APPLICATION TUBING HYDROPHOBIC FILTER PUMP ON OFF SWITCH REGULATOR KNOB VACUUM GAUGE SHORT TUBING 14...

Страница 15: ...15 8 12 Model 1475K 20 COLLECTION JAR LONG TUBING APPLICATION TUBING HYDROPHOBIC FILTER SHORT TUBING...

Страница 16: ...10 REPAIR KITS 9 11 Repair Kit for Model 2515 Service Seal Gasket Description Kit Kit 2515K 01 2515K 02 Connecting Rod Assembly 1 1 Handle Screw 2 2 Head Screw 3 4 Head O Ring 4 1 1 Valve Plate Assem...

Страница 17: ...el Adapter for dispoable tips w o ejector 1475K 03 Rubber Adapter for Pasteur pipettes 1475K 04 8 Channel Adapter for dispoable tips w ejector 1475K 05 Adapter S S 150mm 1475K 06 4 Channel Adapter S S...

Страница 18: ...ription Welch Kit 115V 60Hz Maintained on off footswitch 1430A 115V 60Hz Variable on off footswitch 1430B 90 230V 50 60Hz Variable on off footswitch with IEC connections 1430C 18 Description Welch Kit...

Страница 19: ...19...

Страница 20: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Отзывы: