4. Setting and meaning
To reach the “setting-mode” you have to reboot the scale, so begin by turn it off.
Now hold down the “UNIT”-key and press the “ON/OFF”-key to start the setting
mode. Pressing the “UNIT”-key can change the parameters (C1-C3), pressing
“TARE” can change the values of the chosen parameter (C1---0 to C1---4):
C1---- Sensitivity
setting 0 1 2 3 4, the larger number is, the less sensible but the
better stable.
C2---- Filtering strength
setting 0 1 2 3, the larger number is, the lower reaction
speed but the better the stability of the scale.
C3---- Auto power-off
setting 0 1 2 3 4, 0: always on 1: 15 min. 2: 30 min.
3: 45 min. 4: 60 min. Defaulting setting is 1.
After C3 press again the “UNIT”-key to save the changes and restart the scale in
weigh-mode.