3.7.3

Data format: starting spaces

data

symbol

ending

enter

2unit

7units

3unit

1 unit

1 unit

3.7.4

Calibration: no calibration

3.7.5

Output pins content (9-core socket): 2 pins: RXD

3 pins: TXD

5pins:

GND

3.7.6

Computer can control balance though RS232. The orders to balance are:

1BH+70H: sending data
1BH+71H: calibration (like “CAL”)
1BH+72H:

Counting activated

1BH+73H: Units change
1BH+74H: Tare

3.7.7

software download in our homepage

www.gandg.de

Attachment:

Attachment:

Attachment:

Attachment: settings

settings

settings

settings and

and

and

and meanings

meanings

meanings

meanings

1. Press “CAL” and do not loose till “F----3” (Balance should be switched on)

.

And pressing “CAL” can change the parameters; pressing “TARE” can

change the values of parameters.

2. C1------ sensitivity setting 0 1 2 3 4, the larger number is, the less sensible but

the better stable. Default setting is 2 or 1

3. C2------filtering strength setting 0 1 2 3, the larger number is, the lower

reaction speed, but the better stable. Default setting is 2 or 1

4. C3------Baud rate setting 2(600) 3(1200) 4(2400) 5(4800) 6(9600). Default

setting is 2

5. C4------communication setting. “TARE” changes the lower unit. “COUNT”

changes the higher units. This data is the first data during communication.
Default setting is 27.

6. Re-pressing “CAL” after setting. Then the balance is auto calibrated to 0.

Содержание JJ100Y

Страница 1: ...000Y JJ3000Y JJ5000Y JJ6000Y Operation Manual G G GmbH Koenigstr 34 41460 Neuss Germany G G GmbH Content 1 Overview 2 Specifications and Performance indexes 3 Operational Method 4 Outer structure figure of electronic scale 5 Cautions for operation 6 Windscreen and calibration weights 7 The batteries power socket and RS232 data interface 8 The Warranty 9 Contact ...

Страница 2: ...no other product can compete with the product is already widely used in machinery industry chemical textile as well as food industry where a high standard of measure and analysis procedure is preferred Many companies and enterprises have improved their quality control procedure with lower cost factor and higher working efficiency by adapting our instrument 2 2 2 2 Specifications Specifications Spe...

Страница 3: ... stabilizing press TARE the scale displays zero Then place the article in container at this time it shows the net weight of the article Remove the article and the container The scale shows a negative value of container weight Still press TARE to return to zero 3 5 Count 3 5 1 First choose samples To precisely count the articles firstly the quantity of samples for counting should be chosen in accor...

Страница 4: ...ngs meanings meanings 1 Press CAL and do not loose till F 3 Balance should be switched on And pressing CAL can change the parameters pressing TARE can change the values of parameters 2 C1 sensitivity setting 0 1 2 3 4 the larger number is the less sensible but the better stable Default setting is 2 or 1 3 C2 filtering strength setting 0 1 2 3 the larger number is the lower reaction speed but the b...

Страница 5: ...icles Avoid impact and heavy load on top never exceed its max weighing limit 5 3 Place the scale on a firm table for operation and make sure that the operating environment of the balance is free of great vibration and interference of power supply and of corrosive gases and liquids 5 4 Ensure the warm up time after power on 5 5 Warning if the instrument is used in a manner not specified in this ins...

Страница 6: ...ket socket socket socket and and and and RS232 RS232 RS232 RS232 data data data data interface interface interface interface 7 1 Battery Specification 4 pics C type batteries 7 2 Following is the diagram for battery change 7 3 Here follows the diagram of power socket and RS232 data interface ...

Страница 7: ...ects and warranty cases we are entitled at our own discretion to either replace the delivery at the prices which were valid at the time of the complaint or to rectify the defect 8 3 We have the right to try to replace the defective product at most twice There will be a slight charge for the repair service of those products by which the warranty period expires Defective goods may only be returned t...

Страница 8: ...he groundwater and get into the food chain damaging your health and well being When replacing old appliances with new ones the retailer is legally obligated to take back your old appliance for disposals at least free of charge 9 9 9 9 contact contact contact contact Address G G GmbH Koenigstr 34 41460 Neuss Germany Tel 49 2131 609 7271 Fax 49 2131 609 7273 Email info gandg de Website www gandg de ...

Отзывы: