C

ALIBRATION

 

W

ARNING?

Did you receive a

T

ROUBLESHOOTING

 

A

 

F

AILED

 

C

ALIBRATION

Double-check the following:

•     Regular connections are to 

Calibration

 side of the UDC4.

•     Floating-ground cable is connected to calibration shield.

•     If possible, the 

Chassis Ground

 is attached to a known, good earth ground.

Click the 

Retry

 button, and the rest of the calibration restarts. Click the 

Ignore

 button for 

any other calibration warnings that may appear, and continue to Step 3. 

After a failed calibration attempt:

•  Find 

Calibration Results PC6-#####.txt

 in your 

My Gamry Data

 folder.*

•  E-mail the file, along with complete contact information, to techsupport@gamry.com

#####

 is the serial number of your potentiostat.

Calibration is used to check the potentiostat’s health, and to “zero” many of the measurement circuits to 

your laboratory environment. A warning does not necessarily indicate a critical failure, and Gamry can 

use calibration information to determine the source of the warning.