![Gamry eQCM 10M Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/gamry/eqcm-10m/eqcm-10m_operators-manual_3387510027.webp)
Chapter 5 -- Cell Connections--Quartz Crystal Installation
5 - 3
Figure 5 - 4
Placement of Crystal and Holder on Cell Bottom
Place the main body of the cell over top of the crystal, invert, and use the four supplied screws to hold the
bottom and body of the cell together (see Figure 5.5).
Figure 5 - 5
Assembly of Cell
Do not over tighten or you may damage the threads or the crystal. Lastly, insert the leads from the crystal into
the ceramic adapter on the BNC cable being sure to connect the black lead to the quartz crystal lead below the
black dot on the Teflon cell (see Figure 5.6).