 Shut the toggle valve and uncouple the detector pad assembly.  

z

Unscrew the two halves of the detector pad holder. Using tweezers, take 
the detector pad out of its envelope and put it, orange side up, in the 
recess in the outlet side of the holder. If the sample pad has any 
discoloration, unevenness in dye content, or a faded yellow appearance, 
select another sample pad. Screw the pad holder assembly back together. 
When using a three-way valve, the assembly may be flushed with the pad 
in place. Do not remove the test pad from the hermetically sealed 
envelope until ready for use. The pad can absorb moisture from the air, 
rain, or sneezing. Exposing the test pad to the air on a humid day will 
ruin the pad in a matter of minutes.  

z

Couple the detector pad assembly back to the sampling coupler, with the 
toggle valve closed, and put the end of the plastic tubing into the neck of 
the plastic sample bottle. Open the toggle valve and allow 500 mL of 
product to pass through the sample pad. Normal sample volume is 500 
mL, but if the reading is off scale, sample volumes as low as 100 mL may 
be used.  

z

Close the toggle valve and uncouple the detector pad holder from the 
sampling coupler. Unscrew the detector pad holder. Slip one prong into 
the notch in the pad holder, and lift the pad out. Press the test pad 
between dry paper towels or blotters to remove excess fuel. 

TEST PROCEDURES 

Test the fuel after the equipment has been prepared and the sample is drawn. 
For maximum accuracy, read the test pad within three minutes after sampling 
is begun

.

 Use the following procedures for testing. 

z

Using tweezers, put the pad in the test pad slot in the bottom of the 
ultraviolet lamp assembly. Ensure that the orange side faces the 
ultraviolet lamp. Turn on the lamp. Press the hooded button of the meter 
assembly while moving the light modulator lever as shown in Figure E-8. 
Watch the meter scale while moving the light modulator lever until the 
meter pointer points to zero. Always move the modulator arm in the same 
direction to avoid backlash. Release pressure on the hooded button and 
shut the lamp switch off when the meter pointer has settled on zero. The 
meter pointer should stabilize in about a minute.  

z

Take the reading from the scale below the lever at the point where the 
lever crosses the scale. Record the reading and sample volume. With a 

Page 8 of 10

FM 10-67-1 APPENDIX E

5/30/2005

http://www.globalsecurity.org/military/library/policy/army/fm/10-67-1/APPE.HTML

Содержание Aqua-Glo III Series

Страница 1: ...equal brightness in the ultraviolet light The balance is reached when the zero centering ammeter reads ZERO The amount of water is read in parts per million PPM on the diaphragm lever arm scale that has been precalibrated in the factory Test Pads The fuel sample passes through a uranine dye treated filter pad Free water in the fuel reacts with the uranine dye When the pad is illuminated by ultravi...

Страница 2: ...rom the test pad before testing is conducted for prior to testing for free water Tweezers Tweezers are used at all times when handling the test pad is being handled Figure E 1A Aqua Glo Series III ultraviolet detector kit Figure E 1B Aqua Glo Series III ultraviolet detector kit Page 2 of 10 FM 10 67 1 APPENDIX E 5 30 2005 http www globalsecurity org military library policy army fm 10 67 1 APPE HTM...

Страница 3: ...calibration setting you will use for calibrating The sample pad in Figure E 3 has a calibration set code of 5 3 Ensure that the calibration and fluorescing standard are a matched set The fluorescing standard located under the hinged light shield has an alphanumeric code and should match the code on the calibration standard Although the SET numbers on the standards may or may not be the same the co...

Страница 4: ...bration meter housing using the jeweler s screwdriver Insert the jeweler s screwdriver into the housing and turn the internal adjustment screw Depress the switch button on the instrument pack as you turn the screw until the meter lever reads 0 Replace the outer screw and the water detector is calibrated See Figure E 7 z Remove the calibration pad from the test area using tweezers and place it back...

Страница 5: ...ration pad and calibration set code Figure E 4 Placement of calibration pad in test area window Page 5 of 10 FM 10 67 1 APPENDIX E 5 30 2005 http www globalsecurity org military library policy army fm 10 67 1 APPE HTML ...

Страница 6: ... E 5 Set the light modulator lever Figure E 6 Calibration meter lever shaft reading 0 Page 6 of 10 FM 10 67 1 APPENDIX E 5 30 2005 http www globalsecurity org military library policy army fm 10 67 1 APPE HTML ...

Страница 7: ...toggle valve closed parallel to the line to the sampling coupler The detector pad holder assembly includes plastic tubing the detector pad holder toggle valve and sampling coupler z Flush the sampling line immediately before sampling Put the end of the plastic tubing in a container that will hold approximately 1 gallon Note The amount of free water in a sample is very sensitive to the temperature ...

Страница 8: ...uple the detector pad holder from the sampling coupler Unscrew the detector pad holder Slip one prong into the notch in the pad holder and lift the pad out Press the test pad between dry paper towels or blotters to remove excess fuel TEST PROCEDURES Test the fuel after the equipment has been prepared and the sample is drawn For maximum accuracy read the test pad within three minutes after sampling...

Страница 9: ...hen the meter pointer points to zero and perform the following calculation Free water ppm meter reading ppm 500 sample volume mL If the retest shows more than the 10 PPM of water submit a sample to the supporting laboratory for inspection and disposition instructions and take the fuel and refueler or fuel system out of service CLEANING AND MAINTENANCE Keep the kit and its components clean Store th...

Страница 10: ...Figure J 8 Reading the undissolved water level Page 10 of 10 FM 10 67 1 APPENDIX E 5 30 2005 http www globalsecurity org military library policy army fm 10 67 1 APPE HTML ...

Отзывы: