System Linearity

Ultrasound systems use special processing circuits to translate the amplitude of echoes 
into brightness levels on the video monitor. These circuits use mathematical functions 
that often produce an S-shaped curve when graphed. As shown in the figure below, 
each echo level produces a corresponding brightness level on the monitor.

An S-shaped curve is used to translate echo levels into brightness levels on the video 
display. Notice how each echo level, –X, 0 and +X dB produce the corresponding bright-
ness values B–, Bo and B+.

As long as the shape of the curve remains constant, the contrast, or difference in 
brightness between different echo levels, will remain constant. If the shape of the 
S-curve changes, the relative image brightness for each echo level will also change.
For example, image post-processing techniques help the user identify subtle tissue 
variations by modifying the shape of the S-curve to emphasize certain ranges of echo 
levels. Degradation in the system hardware can also affect the shape of the curve and 
produce unexpected variations in the contrast between echo levels. The distortion in the 
information displayed to the user may affect the interpretation of the ultrasound image. 
This is why it is important to do quality measurement with as little post-processing as 
possible. An example of these distortions is shown on the following page.

Содержание 404GS LE

Страница 1: ...404GS LE Precision Small Parts Grey Scale Phantom Gammex User s Guide...

Страница 2: ...12 Pin Target 12 Resolution Target Group 12 Phantom Specifications 14 Physical Specification 14 Tissue Mimicking Background Material 14 Low Scatter Anechoic Cysts 14 GreyScaleTargets 14 Harmonic Imag...

Страница 3: ...for testing for Axial and lateral resolution targets at 1 3 5 and 6 cm for precise resolution mea surements Low scatter cyst imaging at 1 2 and 4 mm Attenuation coefficients of either 0 5 or 0 7 dB cm...

Страница 4: ...cation Weigh your 404GS LE to monitor desiccation Weigh the phantom when you first receive it and then every 6 months Record the values on the data sheet Do not drop or damage the phantom Return the p...

Страница 5: ...y oil or lanolin based gels as a coupling medium Poor transmission is a result of insufficient coupling Do not press the transducer into the scanning surface This damages the scanning surface and will...

Страница 6: ...l in the 404GS LE is ultrasonically similar to human tissue This allows the use of normal scanner control settings and ensures that the performance measured with the phantom closely approximates the s...

Страница 7: ...ou may notice that the tissue echoes near the bottom of the phantom fade into noise The depth at which usable echoes disappear is called the depth of penetration The depth markers on the phantom label...

Страница 8: ...onic calipers to measure the distance between two of the vertical pin targets Repeat for two of the horizontal pin targets The vertical pins have 5 mm spacing while the horizon tal pins have 10 mm spa...

Страница 9: ...tal width of the pin targets in the near mid and far fields of the image Notice how the pin targets are narrowest in the focal zone The pin width demonstrates the width of the ultrasound beam at that...

Страница 10: ...e normal Measure the dimensions of the 4 mm cystic target to check the image geometry Use the calipers to measure from top to bottom and side to side Repeat with the other cystic target groups as part...

Страница 11: ...cyst lower the gain until it just disappears Record this gain setting and use it for future grey scale measurements Freeze the image and visually evaluate the grey scale targets Match each target wit...

Страница 12: ...eter 7 mm Placement 3 cm deep Pin Targets Diameter 0 1 mm Vertical spacing 5 mm at 1 9 cm deep Horizontalspacing 10mmat1and5cmdeep Resolution Target Groups 1 3 5 and 6 cm deep All acoustic measurement...

Страница 13: ...13...

Страница 14: ...e of speed of sound 1 5 m s C Attenuation coefficient 0 7 0 05 dB cm MHz 0 5 0 05 dB cm MHz refer to phantom side label Low Scatter Anechoic Cysts Speed of sound 1540 10 m s Temperature dependence of...

Страница 15: ...uch attenuation as there would be if the higher frequency were used to generate the pulses at the transducer So higher frequency resolution occurs for greater depths within the subject than if convent...

Страница 16: ...tely the metastases may not be detected The Quantitative Measurement ensures that the grey level signal is measured consistently The Qualitative Measurement ensures that grey levels are displayed on t...

Страница 17: ...nt the contrast or difference in brightness between different echo levels will remain constant If the shape of the S curve changes the relative image brightness for each echo level will also change Fo...

Страница 18: ...ed to B 1 Bo 1 and B 1 Changes in the system response can be identified by measuring the average pixel value of the grey scale targets and the background material as displayed on the video monitor Pix...

Страница 19: ...manner each time Method Define a region of interest and measure the average echo level Procedure Baseline Test 1 Scan the grey scale targets and display them as large as possible Freeze the image 2 M...

Страница 20: ...d by locating a grey bar step that approximates the brightness of the region of interest Note It is absolutely critical that all system control settings be precisely reproduced for these tests Errors...

Страница 21: ...d environments the phantom weight should be checked every six months Phantoms used in high temperature low humidity environments or in mobile situations should be tested more frequently As the phantom...

Страница 22: ...f the defect and d which are found to be defective by Seller upon Seller s examination Buyer shall pay all transportation charges Products manufac tured by third parties must be returned to Seller wit...

Страница 23: ...epartment hours are Monday through Friday 7 30 am to 5 00 pm Central Time 1 800 GAMMEX 1 426 6391 1 608 828 7000 1 608 828 7500 Fax e mail sales gammex com Service Department hours are Monday through...

Страница 24: ...24 Notes...

Страница 25: ...gammex com e mail sales gammex com Gammex RMI LTD Broadway Business Centre 32A Stoney Street Nottingham NG1 1LL United Kingdom 44 0115 9247188 Fax 44 0115 9247189 e mail uksales gammex com Gammex RMI...

Отзывы: