Features

Initialization Sequence

Upon power-on, 163-DIG performs a startup sequence countdown. Keep the unit on a flat and stable 

surface during initialization. At the end of initialization 163-DIG automatically zeroes.

Zero/Tare

Zero/Tare allows users to zero the scale for relative measurements, or measurements in different 

orientations. Before any compression measurement is taken, the 163-DIG should be zeroed.

Note: 

After a tare force is removed, the Scale will read a negative value. Press Zero/Tare to re-zero 

the 163-DIG.

Hold/Pause

Hold/Pause (

) allows users to freeze the display on the Scale. Press it again to cancel, or use 

Zero/Tare.

Overload Warning

Do not exceed the scale capacity – this can cause permanent damage and void warranty. Refer to 

the 

Scale Specifications

 on page 6 for the scales capacity. If the scale is overloaded, the display 

shows Err-0 and the load should be removed immediately. 

Energy Saver

The scale automatically powers off after 90-seconds of inactivity to preserve batteries. Activity 

includes any button presses or change in force measurement.

Table 1: Accessories

Foam Compression Block

Soft Case

 

 

Содержание 163-DIG

Страница 1: ...i Title LJLWDO RPSUHVVLRQ 6FDOH 0RGHO...

Страница 2: ...selling or otherwise distributing any part of this manual without the prior writ ten consent of Gammex Inc is prohibited Gammex Inc reserves the right to make periodic modifications to this manual wit...

Страница 3: ...uence 4 Zero Tare 4 Hold Pause 4 Overload Warning 4 Energy Saver 4 Detachable Display Panel 5 Scale Specifications 6 Using the 163 DIG 7 General Operation 7 Standard Setup 0 degree 7 Prone Stereotacti...

Страница 4: ...type of batteries to use A display of Lo or the battery symbol on the 163 DIG readout indicates low batteries and replacements are required To install batteries in the 163 DIG 1 On the bottom of the 1...

Страница 5: ...t potentially lead to voiding product warranty Independent Failure Mode Warning Do not use results from the 163 DIG to make changes to your system do use the results to perform additional system check...

Страница 6: ...n over time You can freeze the display of the maximum force reading by pressing the Pause Hold button Press the Zero button to reset Components Figure 2 1 Model 163 DIG Diagram Table 2 1 Parts and Acc...

Страница 7: ...ved the Scale will read a negative value Press Zero Tare to re zero the 163 DIG Hold Pause Hold Pause allows users to freeze the display on the Scale Press it again to cancel or use Zero Tare Overload...

Страница 8: ...1 On the bottom push the release tab with arrow indicting direction 2 Wriggle and slide the Display Panel upwards in the direction of the Platform while holding the release tab 3 The separated Displa...

Страница 9: ...g 0 1 kg 0 01 kg 1 25 kg Display Units g lb oz kg lb oz Scale Dimensions 8 9 x 8 2 x 2 9 in 225 x 208 x 73 mm Platform Dimensions 7 87 x 6 22 in 200 x 158 mm Foam Compression Block Dimensions 7 25 x 4...

Страница 10: ...mpression paddle Optionally place the Foam Compression Block on top of the 163 DIG Platform which is also useful for testing curved paddles 3 Power On the 163 DIG Scale and wait for Initialization to...

Страница 11: ...lat and stable surface 2 Rotate the 163 DIG onto its narrow side on a flat and stable surface and zero the scale 3 Bring the 163 DIG to the Mammography system and very carefully apply pressure to clam...

Страница 12: ...mponents are subject to wear Calibration Process Required equipment 10 kg certified weight or combination of certified weights Process 1 While the 163 DIG is off press and hold the Unit button while h...

Страница 13: ...ter use with a damp cloth if necessary Contacting Gammex Support Technical support is available from Gammex Inc Contact the Support department By telephone 1 321 259 6862 option 3 then 3 By email Glob...

Страница 14: ...21 CFR 820 Gammex Quality Management System is certified to 13485 by SGS Operator Responsibility The instructions in this manual are intended for trained personnel The operator is solely responsible f...

Страница 15: ...any safety or health related issues or concerns regarding the use of Gammex products contact Gammex directly Modifications to Equipment WARNING No modifications may be made to this unit without the e...

Страница 16: ...RUSRUDWH HDGTXDUWHUV LVFRYHU ULYH 0LGGOHWRQ 86 VXQQXFOHDU FRP...

Отзывы: