Troubleshooting
69
point.
closed in centric position.
The holding clamps of the writing bow are not firmly
tightened.
Make sure that the holding clamps of the bow are
firmly fixated both on the shaft of the clutch and on
the side arms. Otherwise, additional stress caused
by the mounted styli can cause certain displace-
ments.
The writing bow was not properly adjusted to be
free of tension.
Repeat the procedure of adjusting the writing bow.
Make sure to keep the system free of tension. The
writing bow must not be allowed to move, espe-
cially when removing the axis locators.
After mounting, the system indicates that there is no contact between stylus and flag of either side.
Possible cause of error
Solution
A stylus is not connected.
Plug the stylus into the CADIAX device.
A flag is not connected.
Plug the flag into the CADIAX device.
The right and left measuring sensors are erro-
neously interchanged.
Insert the plugs correctly.
A stylus is outside the allowed measuring area of
the flag due to incorrect mounting.
Repeat mounting the face bow. Make sure that
both styli are positioned to stay on the black
measuring surface of the flag, also during medio-
trusive movements.
A measuring flag is soiled.
Due to the flag being soiled, there is no continuous
electrical contact between the tip of the stylus and
the measuring surface of the flag. Clean the flag
surface according to the provided cleaning instruc-
tions.
A stylus is defective.
Contact an authorized GAMMA service center.
A flag is defective.
Contact an authorized GAMMA service center.
The CADIAX device is defective.
Contact an authorized GAMMA service center.
Recorded curves go "backwards" instead of protrusive.
Possible cause of error
Solution
The right and left measuring sensors are erro-
neously interchanged.
Insert the plugs correctly.
Recorded curves do not begin in the coordinate origin.
Possible cause of error
Solution
The position of the styli has changed between set-
ting the reference position and starting the record-
ing.
Repeat setting the reference position. If necessary,
repeat this procedure before each recording.
Recorded curves show strong artifacts.