Figure 18. TSP pressure window screen
9.17.
TSP advanced options
Available setup:
Pressure window - Invokes pressure window setup.
Control - Sets TSP control in either Amps or Watts.
Filament status clear - Clears filament status for all filaments.
Filament auto advance - Sets filament auto advance option (off, next, rotate).
The ‘Next’ option instructs the TSP to switch to next available filament in
case where active filament burns out, filament status = open?.
The ‘Rotate’ option instructs the TSP to switch to next available filament
following each sublimation cycle.
For example, when degassing TSP filaments, set Filament = 1, Filament auto
advance = Rotate, and Number of cycles = 4. This TSP setup would sublimate
each filament in turn.
Figure 19. TSP advanced options screen
Page 31
900031_J
900031_J - Touchscreen user interface