Y-value is set with screw
C
and can be read di-
rectly on the Vernier scale of the transversal
axis. Positive values mean a movement of the
LJ position to the right side, negative values
mean a shift towards left.
Method B: SCI + protrusive position
With method B, the actual SCI values of the pa-
tient are used for setting the condylar housings.
Determine the arbitrary therapeutic lower jaw
position in mm, using the protrusion value of the
hinge axis for the left and the right side of the
joint.
The value is set with screw
A
– located at the
backside of the housing
– and can be checked
on the Vernier scale of the housing.
The transversal shift (Y) can be adapted analo-
gous to the setting in method A.
After having finished the setting the screws A, B
and C need to be fixed again on both sides.
For a transfer of the new lower jaw position into
the articulator it is necessary to prepare a bite
registration block in the variator. This bite regis-
tration is then used as usual to mount the lower
jaw model in the articulator to the upper jaw.