18
ERROR
WHAT DOES IT
MEAN
WHAT CAN I DO?
Air Pressure Sensor:
The Control Unit
has detected a
problem calibrating
the pneumatic
compression circuit
on start-up.
•
This is most likely to occur if you have restarted
the System with an in
fl
ated Wrap attached.
•
Detach the Wrap and try again.
Self-Test Error –
Air Pump:
The
Control Unit has
detected an electronic
problem in the air
circuit on start-up.
• Disconnect the Wrap from the Control Unit.
• Turn the Control Unit off and on again without
a Wrap connected.
Dry Pump:
The
Control Unit has
detected a dry pump.
In order to prevent
possible damage to
the
fl
uid pump, the
unit will stop therapy.
• Make sure there is water in the ice box.
• Verify that the ice box
fi
lter is not clogged.
• Make sure there are no kinks in the Wrap or
Connector Hose.
• Turn the Control Unit off and on again.
• Disconnect and reconnect the hose from the
Control Unit and the Wrap.
• Prime the Wrap using the following steps:
- Select “No Pressure.”
- Attach the largest Wrap available.
- Lay the Wrap open (not on the body).
- Run the System for 1 minute.
• Prime the System using the following steps:
- Disconnect the hose from the Control Unit.
- Now, look at the Wrap connection location
on the Control Unit. On the top valve,
push the white prong in so that it is even
with the metal connector.
- Make sure you are not fully covering the
opening in the prong.
- Push start, and water should squirt out of
the valve.
- Restart the System.
Over Pressure:
Indicates that the
Control Unit has
exceeded the target
pressure.
• Turn the Control Unit off and back on.
• Apply the same Wrap more tightly.
• Apply a different Wrap.
T R O U B L E S H O O T I N G