Helpful Hints and Recovery Procedures
COBE
®
Spectra™ Apheresis System •
Essentials Guide
11-9
9
Turn the plasma pump counterclockwise approximately
100 revolutions.
10
Close the white pinch clamp on the return line, then disconnect the
return needle.
11
Close the access saline line roller clamp.
12
Remove the disposable tubing set from the front panel:
a
Unlock the centrifuge cover (see “Manually Overriding the
Centrifuge Cover and Door Latches” on page 11-7), then
remove the channel from the filler.
b
Turn the valves to the open/load positions.
c
Press the cartridge clamps to release the pump cartridges. Turn
the pumps counterclockwise and pull the tubing out to unload
the pumps.
d
Remove the lines from the front panel.
e
Discard the containers, bags, and disposable tubing set in the
biohazard disposal container according to your standard
operating procedures.
Resolving Power-Up Tests Failure
If the “Power up tests in progress” appears for more than 60 seconds,
perform one or both of the following:
1
Turn the power switch off and then on again to retry the power-up
tests.
2
Try another power outlet or circuit in case other equipment is
reducing the power available to the Spectra system.
If the problem persists, notify a qualified service technician or
Gambro BCT.
Repositioning the Access Pressure Sensor
Diaphragm
Repositioning the diaphragm with fluid in the disposable tubing set
during Prime or Run mode may be necessary when
•
The access pressure sensor has not been positioned correctly or has
been disengaged from the access pressure sensor housing.
•
The diaphragm is not in the proper position because the pressure
pod seal is dry or cracked.
To reposition the diaphragm in the access pressure sensor so the
transducer in the housing can monitor access pressure accurately,
perform the following:.
1
Remove the access pressure sensor from its housing. Turn
counterclockwise to unlock.
2
If in Run mode,
close
the access line pinch clamp and
open
the
access saline roller clamp.
3
Lower the saline container until the fluid level in the container is
level with the centrifuge cover (see Figure 11-2). Wait
approximately 5 seconds.
4
Raise the saline container until fluid in the container is level with
the pressure sensor (see Figure 11-2).
5
Wait approximately 10 seconds.
6
Load the pressure sensor into its housing, keeping the fluid level in
the saline container level with the sensor. Push in and turn the
sensor clockwise to lock.
7
Rehang the saline container.
Note:
Before repositioning the diaphragm, pause the system.
Содержание Cobe Spectra
Страница 1: ......
Страница 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Страница 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Страница 51: ...COBE Spectra Apheresis System Essentials Guide 3 1 3 Installing Transporting and Returning the Spectra System ...
Страница 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Страница 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Страница 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Страница 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Страница 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Страница 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Страница 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Страница 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Страница 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Страница 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Страница 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Страница 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Страница 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Страница 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...